Ddrcombi Trauma
by Swissray International, Inc.
510(k) K050718
Active
510(k)
Valid until
No expiry
FDA recalls
None ongoingDocuments
Trust
Unverified
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK050718 | Date26 May 2005 | StatusActive |
Recalls
-
- Terminated Z-0697-2019
A component of the device was recalled by a supplier. In rare cases of inadequate maintenance or very high clinical workload, the first steel wire rope of the lifting column, which is designed to take the load, can break without triggering the safety lock. This can lead to overloading of the safety cable, which is not designed to permanently carry the load of the lifting column or continuously move under load. This could cause material fatigue and cause the arm to drop unexpectedly during patient positioning. If the arm drops unexpectedly, patients and users may be seriously injured.
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- 510(k) Number
- FDA Product Code
- MQB
- Device Class
- Class II
- Regulation
- 21 CFR Part 892.1680
- Advisory Committee
- Radiology
- FDA Decision
- SESE Substantially Equivalent