Delfia (Tm) Hfsh Kit
by Lkb Instruments, Inc. 510(k)
Active
510(k) K862043
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK862043 | Date24 Jun 1986 | StatusActive |
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Recalls
-
- Terminated Z-0779-2017
A DELFIA hFSH kit lot failed real time stability testing. The real time stability data demonstrated a decrease in hFSH concentrations between -4.8% to -11.3% for control sera. The -11.3% bias was observed at an hFSH target concentration of 101 IU/L. The change observed in the serum samples within the range 1.76 67.0 IU/L hFSH averaged -2.9%. The reference range of the assay spans 0.60 to 98.0 IU/L.
Specification
- 510(k) Number
- FDA Product Code
- CGJ
- Device Class
- Class I
- Regulation
- 21 CFR Part 862.1300
- Advisory Committee
- Clinical Chemistry
- FDA Decision
- SESE Substantially Equivalent