Deltec Cozmo Insulin Infusion Pump with Cozmonitor Blood Glucose Meter, Models 21-1815, 21-1816, 21-1817
510(k) K062323
Active
Valid until
No expiry
FDA recalls
None ongoingDocuments
Trust
Unverified
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK062323 | Date1 Nov 2006 | StatusActive |
Recalls
-
- Terminated Z-0989-2009
- Terminated Z-0988-2009
Smiths Medical has become aware of a display irregularity with the Deltec Cozmo Insulin Pump, Model 1800 and Model 21-1700 containing Model 1800 software. There have been adverse events reported that the amount of Extended Bolus delivered was not accurately displayed by the Pump. The amount of extended Bolus displayed on the Pump Home Screen 2 and in the Bolus Summary Report is less than what w
-
- Terminated Z-1623-2008
- Terminated Z-1622-2008
Smith Medical discovered an issue with a motor component in specific serial numbers of the Deltec Cozmo Insulin Pump, Model Numbers 1700 and 1800. This issue affects how the motor operates and may cause an over-delivery of insulin which could result in injury to the user.
-
- Terminated Z-1591-2008
Smith Medical has received reports of Deltec Cozmo Insulin Pumps powering down without an alarm. Investigation concluded that if the battery cap is not fully tightened, the Pump may power down and a brief "chirp" will sound. This may occur if the user has not sufficiently tightened the battery cap or if the battery cap is damaged. If a patient is asleep or does not regularly check their blood gl
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- 510(k) Number
- FDA Product Code
- LZG
- Device Class
- Class II
- Regulation
- 21 CFR Part 880.5725
- Advisory Committee
- General Hospital
- FDA Decision
- SESE Substantially Equivalent
Overview
AI-generated summary, 1 Oct 2026
The Deltec Cozmo Insulin Infusion Pump with Cozmonitor Blood Glucose Meter is a Class II medical device designed for continuous subcutaneous insulin infusion. It delivers precise, programmable insulin doses to manage diabetes in hospital and outpatient settings, integrating real-time glucose monitoring via the Cozmonitor module to enhance treatment accuracy.
This device is intended for single-use and requires FDA clearance (510K) under the Special 510(k) review pathway. It is supplied as a reusable pump system with disposable insulin infusion sets and test strips for the Cozmonitor. The pump operates in environments compatible with standard hospital infrastructure and is not MRI-safe. Models 21-1815, 21-1816, and 21-1817 are authorized for use in general hospital settings, as classified under regulation number 880.5725.
Other devices from Smiths Medical MD, Inc.
Questions
AI-generated answers, 1 Oct 2026
What is the Cozmo Insulin Infusion Pump with Cozmonitor designed for?
The Cozmo Insulin Infusion Pump with Cozmonitor is designed for continuous subcutaneous insulin infusion to manage diabetes in both hospital and outpatient settings. It integrates real-time glucose monitoring via the Cozmonitor module to enhance treatment accuracy by providing precise, programmable insulin doses tailored to patient needs.
How is the Cozmo Insulin Infusion Pump supplied, and what components are included?
The Cozmo Insulin Infusion Pump is supplied as a reusable pump system. It requires disposable insulin infusion sets and test strips for the Cozmonitor blood glucose meter. The pump itself is reusable, while the infusion sets and test strips are intended for single-use to ensure sterility and safety.
Is the Cozmo Insulin Infusion Pump sterile, and how should it be handled?
The Cozmo Insulin Infusion Pump is not pre-sterilized as a reusable device, but the disposable components—such as infusion sets and test strips—must be handled under sterile conditions to prevent contamination. Follow manufacturer guidelines for proper cleaning, disinfection, and storage of the reusable pump to maintain its functionality and safety.
Are the models 21-1815, 21-1816, and 21-1817 interchangeable, or do they have specific uses?
The models 21-1815, 21-1816, and 21-1817 are all authorized for use in general hospital settings under regulation number 880.5725. While they share the same core functionality, there may be minor differences in features or software versions. Always verify compatibility with your facility’s specific protocols and consult the manufacturer’s documentation to ensure proper use.
Can the Cozmo Insulin Infusion Pump be used in environments with MRI equipment, and how should it be stored?
The Cozmo Insulin Infusion Pump is not MRI-safe, so it should not be used or stored near MRI machines. For storage, keep the reusable pump in a clean, dry environment away from direct sunlight or extreme temperatures. Follow the manufacturer’s instructions for maintenance, including proper cleaning and calibration, to ensure reliable performance.