- Terminated Z-0825-2015
Two complaints reported that the BIOLOX delta TS Ceramic Femoral Articul/eze12/14 36mm +12 Head (product code 1365-36-740) did not fit like the trial; the definite hip construct felt loose. The loosening refers to tissue tension and not the taper connection between the femoral head and stem. The investigations found that the small size sleeve (+1.5 femoral head offset) was incorrectly used in the