Device HOLDER WITH MAST-CLAMP LONG
Listed
Valid until
Not applicable
FDA recalls
1 ongoingDocuments
None
Trust
Unverified
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | GUDID listing | Number04033817904137 | Date22 Jan 2026 | StatusListed |
Recalls
-
- Ongoing Z-0571-2025
Heart-Lung Machine GUI touchscreen may go dark for 10-15 seconds, safety/performance functions remain operational before reset, control maintained through backup control panel, but second reset may occur, may lead to gas blender standby, then gas blender must be reactivated using user interface on blender, sensors deactivated until new case started; GUI reset can lead to Hypoxia and Hypoperfusion.
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — GUDID listing Listed
Specification
- Trade Name
- Device
- Generic Name
- Cardiopulmonary bypass system, roller type
- Model / Reference
- HOLDER WITH MAST-CLAMP LONG
- Primary DI / UDI-DI
- 04033817904137
- 510(k) Number
- FDA Product Code
- DTQ
- GMDN Term
- Cardiopulmonary bypass system, roller type
- Device Class
- Class II
- Regulation Number
- 870.4220
- GUDID distribution status
- In Commercial Distribution
Other devices from LivaNova Deutschland GmbH
Properties
Other properties (12)
- Sterile: No
- Single Use: No
- Implantable: No
- Contains Latex: No
- Over-the-Counter: No
- Convenience Kit: No
- Combination Product: No
- HCT/P: No
- Premarket Exempt: No
- Lot / Batch Number: No
- Serial Number: No
- Expiration Date: No