Skip to content

Diastat Anti-Mitochondrial Antibody Kit

by Shield Diagnostics, Ltd.

510(k) K921183

Active United States

510(k)

Valid until

No expiry

510(k) clearance is not time-limited

FDA recalls

None ongoing

1 closed

Documents

Trust

Unverified

Registry data, not yet checked by deviCERTS. Source FDA 510(k), synced 22 Oct 2026. What this means

Regulatory status

as of registry sync 22 Oct 2026

RegionAuthorizationNumberDateStatus
United StatesFDA 510(k)SESE Substantially Equivalent · Traditional NumberK921183 Date19 Jun 1992ClearedStatusActive

Recalls

1 recall · FDA

    • Terminated Z-0673-05 FDA Unclassified (pre-2012) initiated 12 Apr 2004

    The data reduction program written for a specific automated microtiter plate instrument (Bio-Tek ELx800 reader) was found to be faulty. This program is used to calculate the cut off values for Diastat Anti-Mitochondrial Antibody Kit. Program should include a borderline specification to indicate repeat testing is needed.

Coverage

1 jurisdiction

United States of America
  • Active
  • Expiring
  • Withdrawn
  • Listed
  • Not covered

United States — FDA 510(k) Active

Specification

510(k)

510(k) Number
FDA Product Code
DBM
Device Class
Class II
Regulation
21 CFR Part 866.5090
Advisory Committee
Immunology
FDA Decision
SESE Substantially Equivalent

Other devices from Shield Diagnostics, Ltd.

See all devices from Shield Diagnostics, Ltd. →