Skip to content

Diastat Anti-Thyroglobulin Kit

by Shield Diagnostics, Ltd.

510(k) K952183

Active United States

510(k)

Valid until

No expiry

510(k) clearance is not time-limited

FDA recalls

None ongoing

1 closed

Documents

Trust

Unverified

Registry data, not yet checked by deviCERTS. Source FDA 510(k), synced 12 Oct 2026. What this means

Regulatory status

as of registry sync 12 Oct 2026

RegionAuthorizationNumberDateStatus
United StatesFDA 510(k)SESE Substantially Equivalent · Traditional NumberK952183 Date30 Jun 1995ClearedStatusActive

Recalls

1 recall · FDA

    • Terminated Z-0675-05 FDA Unclassified (pre-2012) initiated 12 Apr 2004

    The data reduction program written for a specific automated microtiter plate reader (Bio-Tek ELx800) was found to be faulty. This program is used to calculate the cut off values for the Diastat Anti-Thyroglobulin Kit assay. The negative Control Upper Limit should be less than 0.95, rather than less than 1.0.

Coverage

1 jurisdiction

United States of America
  • Active
  • Expiring
  • Withdrawn
  • Listed
  • Not covered

United States — FDA 510(k) Active

Specification

510(k)

510(k) Number
FDA Product Code
JZO
Device Class
Class II
Regulation
21 CFR Part 866.5870
Advisory Committee
Immunology
FDA Decision
SESE Substantially Equivalent

Other devices from Shield Diagnostics, Ltd.

See all devices from Shield Diagnostics, Ltd. →