DigitalDiagnost C50
by Philips Healthcare (Suzhou) Co., Ltd.
Listed
Valid until
Not applicable
FDA recalls
3 ongoingDocuments
None
Trust
Unverified
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | GUDID listing | Number00884838082366 | Date26 Oct 2017 | StatusListed |
Recalls
-
- Ongoing Z-1383-2023
- Ongoing Z-1382-2023
If the PM for the telescopic carriage is not performed annually as defined in the Instructions for Use (IFU), the tension force of the spring balancer inside of the telescopic carriage will unavoidably decay through normal wear and tear; as a result there is a potential that the Ceiling Suspension of the DigitalDiagnost C50 system may descend inadvertently if power to the system is interrupted, and may result in injuries to patients and users
-
- Ongoing Z-1563-2022
Following a system restart where the Table Height 2 (TH2) is the default setting and then switched to Wallstand VS2, operators will see an incorrect orientation of the image on the first examination due to an issue in the firmware of the Wallstand VS2 board. The system will rotate the amplimat field selection by 90 degrees. The wrong amplimat field selection may cause an incorrect dose of radiation to occur. Additionally, the anatomic position markers may become mispositioned and could potentially be associated with the opposite side of the anatomy. (Added 1/3/2023), Expansion of the root cause to include incorrect amplimat field selection by 90 degrees following the first exposure using Wallstand VS2 and a subsequent system restart. Previously, this issue only occurred following a system restart where the Table Height 2 (TH2) is the default setting.
-
- Terminated Z-2498-2019
During recent evaluations of the Philips DigitalDiagnost C50 system, Philips identified a missing label. This notification is intended to provide you with information regarding what the issue is, and relevant affected product, the actions you are required to take and the actions planned by Philips to correct the issue. If you need any or further information or support concerning this issue, please contact your local Philips representative or Philips Customer Service: 1-800-722-9377.
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — GUDID listing Listed
Specification
- Generic Name
- Stationary basic diagnostic x-ray system, digital
- Model / Ref / Catalog
- DigitalDiagnost C50, 712201
- Primary DI / UDI-DI
- 00884838082366
- 510(k) Number
- FDA Product Code
- KPR
- GMDN Term
- Stationary basic diagnostic x-ray system, digital
- Device Class
- Class II
- Regulation Number
- 892.1680
- GUDID distribution status
- Not in Commercial Distribution
Other devices from Philips Healthcare (Suzhou) Co., Ltd.
Properties
Other properties (12)
- Sterile: No
- Single Use: No
- Implantable: No
- Contains Latex: No
- Over-the-Counter: No
- Convenience Kit: No
- Combination Product: No
- HCT/P: No
- Premarket Exempt: No
- Lot / Batch Number: No
- Expiration Date: No
- Manufacturing Date: No