DigitalDiagnost C50

by Philips Healthcare (Suzhou) Co., Ltd.

Listed FDA GUDID listing, not in commercial distribution

Valid until

Not applicable

Database listing, no authorization

FDA recalls

3 ongoing

This device is named in the recall

Documents

None

No authorization documents

Trust

Unverified

Registry data, not yet checked by deviCERTS. Source GUDID, synced 21 Oct 2026. What this means

Regulatory status

as of registry sync 21 Oct 2026

RegionAuthorizationNumberDateStatus
United StatesGUDID listing Number00884838082366 UDI-DI Date26 Oct 2017ListedStatusListed GUDID: not in commercial distribution

Recalls

4 recalls · 3 ongoing · FDA

  • 2 recalls, same reason

    • Ongoing Z-1383-2023 FDA Class II initiated 9 Mar 2023
    • Ongoing Z-1382-2023 FDA Class II initiated 9 Mar 2023

    If the PM for the telescopic carriage is not performed annually as defined in the Instructions for Use (IFU), the tension force of the spring balancer inside of the telescopic carriage will unavoidably decay through normal wear and tear; as a result there is a potential that the Ceiling Suspension of the DigitalDiagnost C50 system may descend inadvertently if power to the system is interrupted, and may result in injuries to patients and users

    • Ongoing Z-1563-2022 FDA Class II initiated 25 Feb 2022

    Following a system restart where the Table Height 2 (TH2) is the default setting and then switched to Wallstand VS2, operators will see an incorrect orientation of the image on the first examination due to an issue in the firmware of the Wallstand VS2 board. The system will rotate the amplimat field selection by 90 degrees. The wrong amplimat field selection may cause an incorrect dose of radiation to occur. Additionally, the anatomic position markers may become mispositioned and could potentially be associated with the opposite side of the anatomy. (Added 1/3/2023), Expansion of the root cause to include incorrect amplimat field selection by 90 degrees following the first exposure using Wallstand VS2 and a subsequent system restart. Previously, this issue only occurred following a system restart where the Table Height 2 (TH2) is the default setting.

    • Terminated Z-2498-2019 FDA Class II initiated 21 Dec 2018

    During recent evaluations of the Philips DigitalDiagnost C50 system, Philips identified a missing label. This notification is intended to provide you with information regarding what the issue is, and relevant affected product, the actions you are required to take and the actions planned by Philips to correct the issue. If you need any or further information or support concerning this issue, please contact your local Philips representative or Philips Customer Service: 1-800-722-9377.

Coverage

1 jurisdiction

United States of America
  • Active
  • Expiring
  • Withdrawn
  • Listed
  • Not covered

United States — GUDID listing Listed

Specification

GUDID

Generic Name
Stationary basic diagnostic x-ray system, digital
Model / Ref / Catalog
DigitalDiagnost C50, 712201
Primary DI / UDI-DI
00884838082366
510(k) Number
FDA Product Code
KPR
GMDN Term
Stationary basic diagnostic x-ray system, digital
Device Class
Class II
Regulation Number
892.1680
GUDID distribution status
Not in Commercial Distribution

Other devices from Philips Healthcare (Suzhou) Co., Ltd.

Properties

  • Prescription Use
  • Serial Number

Other properties (12)

  • Sterile: No
  • Single Use: No
  • Implantable: No
  • Contains Latex: No
  • Over-the-Counter: No
  • Convenience Kit: No
  • Combination Product: No
  • HCT/P: No
  • Premarket Exempt: No
  • Lot / Batch Number: No
  • Expiration Date: No
  • Manufacturing Date: No