Skip to content

Digora Optime (or Digora Ontime)

by Soredex Palodex Group OY 510(k)

Active United States

510(k) K041050

Regulatory status

as of 29 Oct 2026

RegionAuthorizationNumberDateStatus
United StatesFDA 510(k)SESE Substantially Equivalent · Traditional NumberK041050 Date15 Jun 2004ClearedStatusActive

Coverage

1 jurisdiction

United States of America
  • Active
  • Expiring
  • Withdrawn
  • Listed
  • Not covered

United States — FDA 510(k) Active

Recalls

4 recalls · FDA

  • 4 recalls, same reason

    • Terminated Z-0839-2020 FDA Class II initiated 7 Jan 2020 on shared 510(k) K041050
    • Terminated Z-0841-2020 FDA Class II initiated 7 Jan 2020 on shared 510(k) K041050
    • Terminated Z-0842-2020 FDA Class II initiated 7 Jan 2020 on shared 510(k) K041050
    • Terminated Z-0840-2020 FDA Class II initiated 7 Jan 2020 on shared 510(k) K041050

    Slow degradation of plastic insulation used in the PSU DC plug of the DIGORA Optime, Scan eXam, Express, GXPS-500 systems may result in short circuiting which may cause heating and melting. The melting may generate smoke and eventually cause the PSU to fail.

Recalls on a shared submission name another device cleared under the same FDA submission as this one.

Specification

510(k)

510(k) Number
FDA Product Code
MUH
Device Class
Class II
Regulation
21 CFR Part 872.1800
Advisory Committee
Dental
FDA Decision
SESE Substantially Equivalent

Other devices from Soredex Palodex Group OY

See all devices from Soredex Palodex Group OY →