Dimension Enzymatic Creatinine Flex Reagent Cartridge DF270
510(k) K073055
Active
Valid until
No expiry
FDA recalls
None ongoingDocuments
Trust
Unverified
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK073055 | Date3 Jan 2008 | StatusActive |
Recalls
-
- Terminated Z-2284-2020
The presence of N-acetyl-p-benzoquinone imine (NAPQI), a metabolite of acetaminophen, may cause interference in certain chemistry assays when testing patient samples.
-
- Terminated Z-1232-2020
There is a potential for falsely depressed creatinine results for patients on phenindione therapy when using the enzymatic methodology. Interference has not been observed with the creatinine Jaffe methodology.
-
- Terminated Z-1660-2016
Siemens Healthcare Diagnostics has become aware of N-Acetylcysteine (NAC) and Metamizole (Dipyrone) interference with Trinder and Trinder-like reaction assays. Siemens has confirmed that falsely depressed results may occur on samples drawn from patients receiving N-Acetylcysteine (NAC) or Metamizole. NAC is the accepted antidote for acetaminophen toxicity and is justified in patients at signific
-
- Terminated Z-2177-2011
Patient results may be falsely depressed or falsely elevated across the assay range.
-
- Terminated Z-0485-2009
Reagent may exhibit unflagged inaccurate patient sample results.
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- Full name
- Dimension Enzymatic Creatinine Flex Reagent Cartridge with Model DF270
- 510(k) Number
- FDA Product Code
- JFY
- Device Class
- Class II
- Regulation
- 21 CFR Part 862.1225
- Advisory Committee
- Clinical Chemistry
- FDA Decision
- SESE Substantially Equivalent
Overview
AI-generated summary, 1 Oct 2026
The Dimension Enzymatic Creatinine Flex Reagent Cartridge DF270 is an enzymatic method in-vitro diagnostic device designed for the quantitative measurement of creatinine in human serum, plasma, or urine samples. It employs a biochemical reaction involving creatinase, sarcosine oxidase, and horseradish peroxidase to generate a measurable signal, enabling accurate creatinine assessment in clinical chemistry workflows.
This single-use reagent cartridge is intended for use exclusively on Dade Behring Dimension clinical chemistry systems. It is supplied as a sterile, disposable component and requires no additional sterilization. The device is classified as Class II under FDA regulation 862.1225 and has received FDA 510(k) clearance (K073055) as substantially equivalent. Storage conditions and compatibility with the Dimension system must be followed to ensure accurate results.
Other devices from Dade Behring, Inc.
Questions
AI-generated answers, 1 Oct 2026
What is the Dimension Enzymatic Creatinine Flex Reagent Cartridge DF270 used for?
The DF270 is designed for the quantitative measurement of creatinine in human serum, plasma, or urine samples. It uses an enzymatic method involving creatinase, sarcosine oxidase, and horseradish peroxidase to generate a measurable signal, enabling accurate assessment of kidney function in clinical chemistry workflows.
How is the DF270 supplied and what are its sterility and reuse requirements?
The DF270 is supplied as a single-use, sterile, disposable reagent cartridge. It requires no additional sterilization and is intended for exclusive use on Dade Behring Dimension clinical chemistry systems. Due to its single-use nature, it cannot be reused or reprocessed.
Is the DF270 compatible with all Dade Behring Dimension systems?
No, the DF270 is designed specifically for use on Dade Behring Dimension clinical chemistry systems. While it is compatible with these systems, it is critical to follow the manufacturer’s guidelines and system-specific instructions to ensure accurate and reliable results.
What storage conditions should be followed for the DF270?
The DF270 must be stored according to the manufacturer’s instructions to maintain its integrity and performance. While the exact storage conditions are not detailed in the provided data, it is generally recommended to store reagent cartridges in a cool, dry place, protected from direct light and extreme temperatures, as per the accompanying documentation.
What regulatory clearance does the DF270 hold?
The DF270 has received FDA 510(k) clearance (K073055) and is classified as a Class II device under FDA regulation 862.1225. It is determined to be substantially equivalent to a predicate device, ensuring it meets the necessary safety and performance standards for clinical use.