Dimension EXL High-Sensitivity Troponin I (TNIH) Assay
by Siemens Healthcare Diagnostics, Inc.
510(k) K190675
Active
510(k)
Valid until
No expiry
FDA recalls
None ongoingDocuments
Trust
Unverified
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK190675 | Date17 May 2019 | StatusActive |
Recalls
-
- Terminated Z-0052-2021
Positive bias across the Analytical Measurement Range of the TNIH assay compared to an unaffected control lot. The average bias observed for patient samples using lots EB0255 and EA0178 of the Dimension EXL Troponin compared with a control lot was +25%. A maximum bias of +34% in patient samples around the 99th percentile observed
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- 510(k) Number
- FDA Product Code
- MMI
- Device Class
- Class II
- Regulation
- 21 CFR Part 862.1215
- Advisory Committee
- Clinical Chemistry
- FDA Decision
- SESE Substantially Equivalent