Dimension Exl with Lm Clinical Chemistry System, Dimension FT4L Flex Reagent Cartridge, and Loci Thyroid Calibrator, Mod

by Dade Behring, Inc.

510(k) K073604

Active United States

510(k)

Valid until

No expiry

510(k) clearance is not time-limited

FDA recalls

None ongoing

7 closed

Documents

Trust

Unverified

Registry data, not yet checked by deviCERTS. Source FDA 510(k), synced 9 Oct 2026. What this means

Regulatory status

as of registry sync 9 Oct 2026

RegionAuthorizationNumberDateStatus
United StatesFDA 510(k)SESE Substantially Equivalent · Traditional NumberK073604 Date13 Mar 2008ClearedStatusActive

Recalls

7 recalls · FDA

    • Terminated Z-1615-2015 FDA Class II initiated 2 Apr 2015 on shared 510(k) K073604

    Certain Dimension Integrated Chemistry Systems Instrument Power Configurations may have been installed with a non-standard instrument power plug.The use of a non-standard plug configuration as described may lead to a potential electrical hazard. The likelihood of potential clinical impact or injury due to this issue is extremely low.

    • Terminated Z-2133-2014 FDA Class III initiated 27 May 2014 on shared 510(k) K073604

    There is a misassembly issue with the photometer installed on Dimension Integrated Chemistry Systems instruments. Through internal investigation by Siemens Healthcare Diagnostics has confirmed that a limited number of Dimension photometers were misassembled. These photometers are performing within specifications, but may lead to a shortened life of the source lamp.

  • 3 recalls, same reason

    • Terminated Z-1048-2013 FDA Class III initiated 27 Feb 2013 on shared 510(k) K073604
    • Terminated Z-1046-2013 FDA Class III initiated 27 Feb 2013 on shared 510(k) K073604
    • Terminated Z-1047-2013 FDA Class III initiated 27 Feb 2013 on shared 510(k) K073604

    Through investigation of customer complaints, Siemens Healthcare Diagnostics has confirmed an increase in System Check failures due to high System Check LOCI results. Upon further investigation, Siemens has determined that many of these errors are due to the current limit set for System Check LOCI results. The errors have no impact on results, but they may cause customer inconvenience. Siemens h

    • Terminated Z-0458-2011 FDA Unclassified (pre-2012) initiated 12 Apr 2010

    Some LOCI cartridges have sub-optimal seal of the lidstock that can cause rupture of the package at high elevations, causing the cartridge to leak. Use of leaking cartridges may result in inaccurate test results.

    • Terminated Z-0581-2009 FDA Unclassified (pre-2012) initiated 30 Sep 2008 on shared 510(k) K073604

    Error messages, "Wash Aspirator Failure" and "Wash station unable to recover" are not generated when one or both HM wash probes are totally occluded.

Recalls on a shared submission name another device cleared under the same FDA submission as this one.

Coverage

1 jurisdiction

United States of America
  • Active
  • Expiring
  • Withdrawn
  • Listed
  • Not covered

United States — FDA 510(k) Active

Specification

510(k)

510(k) Number
FDA Product Code
CEC
Device Class
Class II
Regulation
21 CFR Part 862.1695
Advisory Committee
Clinical Chemistry
FDA Decision
SESE Substantially Equivalent