Dimension Iron Calibrator

by Dade Behring, Inc. 510(k)

Active United States

510(k) K060266

Regulatory status

as of 28 Oct 2026

RegionAuthorizationNumberDateStatus
United StatesFDA 510(k)SESE Substantially Equivalent · Traditional NumberK060266 Date9 Mar 2006ClearedStatusActive

Coverage

1 jurisdiction

United States of America
  • Active
  • Expiring
  • Withdrawn
  • Listed
  • Not covered

United States — FDA 510(k) Active

Recalls

1 recall · FDA

    • Terminated Z-0745-2013 FDA Class III initiated 20 Dec 2012

    Siemens Healthcare Diagnostics has received complaints of IRON calibrations outside of acceptable calibration slope (m) values (0.97 - 1.03) when using IRON calibrator lots 2DN060 and 2FN063. Lots 2GN068 and 2JN074 may also be impacted but no complaints have been received.

Specification

510(k)

510(k) Number
FDA Product Code
JIS
Device Class
Class II
Regulation
21 CFR Part 862.1150
Advisory Committee
Clinical Chemistry
FDA Decision
SESE Substantially Equivalent