Dimension (Tm) Clinical Chemistry System
by E.I. Dupont DE Nemours & Co., Inc.
510(k) K861700
Active
Valid until
No expiry
FDA recalls
1 ongoingDocuments
Trust
Unverified
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK861700 | Date11 Jun 1986 | StatusActive |
Recalls
-
- Ongoing Z-1816-2022
Dimension Magnesium (MG) Flex reagent cartridge lots may exhibit imprecision for Quality Control (QC) and produce Abnormal Reaction flags. The issue is not always detected by QC and erroneous results may be produced in the absence of an Abnormal Reaction flag. Based on customer data, imprecision leading to erroneous, unflagged patient results with a bias of -15 to -59% may occur
-
- Terminated Z-1049-2013
Siemens Healthcare Diagnostics has initiated a voluntary field action for Dimension(R) PHOS Flex(R) Reagent due to reagent interaction causing falsely depressed Creatinine results when EZCR is processed from open wells that are in close proximity to open wells of PHOS reagent.
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- 510(k) Number
- FDA Product Code
- CHH
- Device Class
- Class I
- Regulation
- 21 CFR Part 862.1175
- Advisory Committee
- Clinical Chemistry
- FDA Decision
- SESE Substantially Equivalent