Dimension Tni Flex Reagent Cartridge, and Ctni Sample Diluent with Models, RF621, KD692
by Siemens Healthcare Diagnostics
510(k) K081643
Active
Valid until
No expiry
FDA recalls
None ongoingDocuments
Trust
Unverified
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK081643 | Date3 Jul 2008 | StatusActive |
Recalls
-
- Terminated Z-2436-2018
The diluent may have an incomplete slit on the septum in the cap of the vial. A diluent cap with a defective septum has the potential to cause erroneously elevated results if a sample is autodiluted onboard the system, due to a reduced volume of diluent being pipetted from the vial.
-
- Terminated Z-2606-2017
The concentrations for Biotin listed in the Non-Interfering Substances section of the current Dimension and Dimension Vistat Instructions for Use (IFU) for Dimension TNI, Dimension Vista DIGXN, E2, FERR, PRL, and TSH have incorrect units, and incorrectly state the level at which biotin does not interfere.
-
- Terminated Z-0069-2012
Firm has confirmed customer complaints for low recovery of QC and patient samples.
-
- Terminated Z-0460-2011
Some LOCI cartridges have sub-optimal seal of the lidstock that can cause rupture of the package at high elevations, causing the cartridge to leak. Use of leaking cartridges may result in inaccurate test results.
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- 510(k) Number
- FDA Product Code
- MMI
- Device Class
- Class II
- Regulation
- 21 CFR Part 862.1215
- Advisory Committee
- Clinical Chemistry
- FDA Decision
- SESE Substantially Equivalent