Dimension Tni Flex Reagent Cartridge, and Ctni Sample Diluent with Models, RF621, KD692

by Siemens Healthcare Diagnostics

510(k) K081643

Active United States

510(k)

Valid until

No expiry

510(k) clearance is not time-limited

FDA recalls

None ongoing

4 closed

Documents

Trust

Unverified

Registry data, not yet checked by deviCERTS. Source FDA 510(k), synced 22 Oct 2026. What this means

Regulatory status

as of registry sync 22 Oct 2026

RegionAuthorizationNumberDateStatus
United StatesFDA 510(k)SESE Substantially Equivalent · Special NumberK081643 Date3 Jul 2008ClearedStatusActive

Recalls

4 recalls · FDA

    • Terminated Z-2436-2018 FDA Class II initiated 21 Mar 2018

    The diluent may have an incomplete slit on the septum in the cap of the vial. A diluent cap with a defective septum has the potential to cause erroneously elevated results if a sample is autodiluted onboard the system, due to a reduced volume of diluent being pipetted from the vial.

    • Terminated Z-2606-2017 FDA Class II initiated 21 Mar 2017

    The concentrations for Biotin listed in the Non-Interfering Substances section of the current Dimension and Dimension Vistat Instructions for Use (IFU) for Dimension TNI, Dimension Vista DIGXN, E2, FERR, PRL, and TSH have incorrect units, and incorrectly state the level at which biotin does not interfere.

    • Terminated Z-0069-2012 FDA Unclassified (pre-2012) initiated 31 Aug 2011 on shared 510(k) K081643

    Firm has confirmed customer complaints for low recovery of QC and patient samples.

    • Terminated Z-0460-2011 FDA Unclassified (pre-2012) initiated 12 Apr 2010

    Some LOCI cartridges have sub-optimal seal of the lidstock that can cause rupture of the package at high elevations, causing the cartridge to leak. Use of leaking cartridges may result in inaccurate test results.

Recalls on a shared submission name another device cleared under the same FDA submission as this one.

Coverage

1 jurisdiction

United States of America
  • Active
  • Expiring
  • Withdrawn
  • Listed
  • Not covered

United States — FDA 510(k) Active

Specification

510(k)

510(k) Number
FDA Product Code
MMI
Device Class
Class II
Regulation
21 CFR Part 862.1215
Advisory Committee
Clinical Chemistry
FDA Decision
SESE Substantially Equivalent