Dimension Visa HBA1C, Model K3105A

by Siemens Heatlhcare Diagnostics, Inc.

510(k) K102045

Active United States

510(k)

Valid until

No expiry

510(k) clearance is not time-limited

FDA recalls

None ongoing

3 closed

Documents

Trust

Unverified

Registry data, not yet checked by deviCERTS. Source FDA 510(k), synced 22 Oct 2026. What this means

Regulatory status

as of registry sync 22 Oct 2026

RegionAuthorizationNumberDateStatus
United StatesFDA 510(k)SESE Substantially Equivalent · Traditional NumberK102045 Date14 Feb 2011ClearedStatusActive

Recalls

3 recalls · FDA

    • Terminated Z-0420-2015 FDA Class II initiated 30 Sep 2014

    Certain lots of the reagent cartridges show confirmed positive bias averaging 0.4% Hemoglobin A1c units and occasionally up to 1.0% HbA1c units for patient samples when compared to the National Glycohemoglobin Standardization Program (NGSP). QC samples may exhibit a similar bias.

    • Terminated Z-0127-2015 FDA Class II initiated 14 Aug 2014

    In March, Siemens Healthcare Diagnostics issued an Urgent Medical Device Correction for decreased recovery of patient samples with the Dimension Vista® CSA & CSAE Flex® reagent cartridges when whole blood samples are run from Small Sample Containers (SSC). Siemens has confirmed that when whole blood samples are run from SSC, HbA1c values may exhibit a percent bias ranging from - 8.2% t

    • Terminated Z-0626-2014 FDA Class II initiated 22 Apr 2013

    Positive bias on the lots listed compared to alternate methods and a higher frequency of "Above Assay Range" flags with the lots listed.

Coverage

1 jurisdiction

United States of America
  • Active
  • Expiring
  • Withdrawn
  • Listed
  • Not covered

United States — FDA 510(k) Active

Specification

510(k)

510(k) Number
FDA Product Code
LCP
Device Class
Class II
Regulation
21 CFR Part 864.7470
Advisory Committee
Hematology
FDA Decision
SESE Substantially Equivalent

Other devices from Siemens Heatlhcare Diagnostics, Inc.