Dimension Visa HBA1C, Model K3105A
by Siemens Heatlhcare Diagnostics, Inc.
510(k) K102045
Active
Valid until
No expiry
FDA recalls
None ongoingDocuments
Trust
Unverified
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK102045 | Date14 Feb 2011 | StatusActive |
Recalls
-
- Terminated Z-0420-2015
Certain lots of the reagent cartridges show confirmed positive bias averaging 0.4% Hemoglobin A1c units and occasionally up to 1.0% HbA1c units for patient samples when compared to the National Glycohemoglobin Standardization Program (NGSP). QC samples may exhibit a similar bias.
-
- Terminated Z-0127-2015
In March, Siemens Healthcare Diagnostics issued an Urgent Medical Device Correction for decreased recovery of patient samples with the Dimension Vista® CSA & CSAE Flex® reagent cartridges when whole blood samples are run from Small Sample Containers (SSC). Siemens has confirmed that when whole blood samples are run from SSC, HbA1c values may exhibit a percent bias ranging from - 8.2% t
-
- Terminated Z-0626-2014
Positive bias on the lots listed compared to alternate methods and a higher frequency of "Above Assay Range" flags with the lots listed.
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- 510(k) Number
- FDA Product Code
- LCP
- Device Class
- Class II
- Regulation
- 21 CFR Part 864.7470
- Advisory Committee
- Hematology
- FDA Decision
- SESE Substantially Equivalent