Skip to content

Dimension Vista Carbon Dioxide, Creatinine and Lactate Dehydrogenase Flex Reagent Cartridge

by Dade Behring, Inc.

510(k) K061238

Active United States

510(k)

Valid until

No expiry

510(k) clearance is not time-limited

FDA recalls

None ongoing

1 closed

Documents

Trust

Unverified

Registry data, not yet checked by deviCERTS. Source FDA 510(k), synced 22 Oct 2026. What this means

Regulatory status

as of registry sync 22 Oct 2026

RegionAuthorizationNumberDateStatus
United StatesFDA 510(k)SESE Substantially Equivalent · Special NumberK061238 Date25 May 2006ClearedStatusActive

Recalls

1 recall · FDA

    • Terminated Z-2741-2015 FDA Class II initiated 15 Jul 2015 on shared 510(k) K061238

    Siemens has confirmed that the Dimension CREA and Dimension Vista CREA may produce falsely depressed results up to - 12.7% and -22% respectively, for a creatinine concentration of 1.5 mg/dL when hemolysis exceeds 200 mg/dL and 300 mg/dL respectively.

Recalls on a shared submission name another device cleared under the same FDA submission as this one.

Coverage

1 jurisdiction

United States of America
  • Active
  • Expiring
  • Withdrawn
  • Listed
  • Not covered

United States — FDA 510(k) Active

Specification

510(k)

510(k) Number
FDA Product Code
KHS
Device Class
Class II
Regulation
21 CFR Part 862.1160
Advisory Committee
Clinical Chemistry
FDA Decision
SESE Substantially Equivalent

Other devices from Dade Behring, Inc.

See all devices from Dade Behring, Inc. →