Dimension Vista Chemistry 1 Calibrator, Model KC110
510(k) K061838
Active
Valid until
No expiry
FDA recalls
None ongoingDocuments
Trust
Unverified
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK061838 | Date21 Aug 2006 | StatusActive |
Recalls
-
- Terminated Z-0652-2017
Siemens determined that discrepant low calcium recovery of QC and patient samples can occur with Dimension Vista CHEM 1 CAL lots 5GM081 and 5GM082.
-
- Terminated Z-2720-2014
Siemens Healthcare Diagnostics has received complaints for low recovery of Magnesium (MG) Quality Control (QC) and patient samples following calibration of MG with the Dimension Vista® CHEM 1 CAL, lot 3GM081. Investigation by Siemens confirms that QC and patient samples under-recovers MG by 0.25 mg/dL [0.10 mmol/L] across the assay range following calibration with 3GM081.
-
- Terminated Z-0864-2010
Siemens Healthcare Diagnostics, Inc. has confirmed high accuracy shifts with Dimension Vista URCA Flex reagent cartridge (various lots) when calibrating with some vials from CHEM 1 calibrator lot 9DM001.
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- 510(k) Number
- FDA Product Code
- JIX
- Device Class
- Class II
- Regulation
- 21 CFR Part 862.1150
- Advisory Committee
- Clinical Chemistry
- FDA Decision
- SESE Substantially Equivalent