Dimension Vista Enzymatic Creatinine Flex Reagent Cartridge, and Enzymatic Creatinine Calibrator, Models K1270, KC270

by Siemens Healthcare Diagnostics, Inc.

510(k) K090330

Active United States

510(k)

Valid until

No expiry

510(k) clearance is not time-limited

FDA recalls

None ongoing

5 closed

Documents

Trust

Unverified

Registry data, not yet checked by deviCERTS. Source FDA 510(k), synced 10 Oct 2026. What this means

Regulatory status

as of registry sync 10 Oct 2026

RegionAuthorizationNumberDateStatus
United StatesFDA 510(k)SESE Substantially Equivalent · Traditional NumberK090330 Date22 May 2009ClearedStatusActive

Recalls

5 recalls · FDA

    • Terminated Z-2285-2020 FDA Class II initiated 4 May 2020

    The presence of N-acetyl-p-benzoquinone imine (NAPQI), a metabolite of acetaminophen, may cause interference in certain chemistry assays when testing patient samples.

    • Terminated Z-1233-2020 FDA Class II initiated 18 Dec 2019

    There is a potential for falsely depressed creatinine results for patients on phenindione therapy when using the enzymatic methodology. Interference has not been observed with the creatinine Jaffe methodology.

    • Terminated Z-1664-2016 FDA Class II initiated 17 Mar 2016

    Siemens Healthcare Diagnostics has become aware of N-Acetylcysteine (NAC) and Metamizole (Dipyrone) interference with Trinder and Trinder-like reaction assays. Siemens has confirmed that falsely depressed results may occur on samples drawn from patients receiving N-Acetylcysteine (NAC) or Metamizole. NAC is the accepted antidote for acetaminophen toxicity and is justified in patients at signific

    • Terminated Z-2154-2011 FDA Unclassified (pre-2012) initiated 10 Jun 2010

    Results may be falsely depressed up to 0.4 mg/dL or falsely elevated up to 0.6 mg/dL across the assay range.

    • Terminated Z-2419-2010 FDA Unclassified (pre-2012) initiated 12 Feb 2010

    The ECREA test results may be falsely elevated or depressed.

Coverage

1 jurisdiction

United States of America
  • Active
  • Expiring
  • Withdrawn
  • Listed
  • Not covered

United States — FDA 510(k) Active

Specification

510(k)

510(k) Number
FDA Product Code
JFY
Device Class
Class II
Regulation
21 CFR Part 862.1225
Advisory Committee
Clinical Chemistry
FDA Decision
SESE Substantially Equivalent