Dimension Vista® CHEM 1 CAL 10716281
by Siemens Healthcare Diagnostics Inc.
Listed
Valid until
Not applicable
FDA recalls
2 ongoingDocuments
None
Trust
Unverified
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | GUDID listing | Number00630414948508 | Date24 Sep 2016 | StatusListed |
Recalls
-
- Ongoing Z-0767-2025
Siemens Healthcare Diagnostics is recalling two lots of their Dimension Vista Total Magnesium (MG) Flex Reagent Cartridges (lots 24057BB and 24064BC) due to the potential for falsely low measurements of magnesium in plasma. Because the error is intermittent, existing quality control procedures may not recognize that a patients results are incorrect, and the results may be reported out
-
- Ongoing Z-0830-2022
Dimension Vista Magnesium Flex reagent cartridge lots exhibiting imprecision and intermittent Abnormal Assay flags. When this issue occurs, there is a potential for QC failures or erroneous patient results which may lead to repeat and follow-up testing.
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — GUDID listing Listed
Specification
- Trade Name
- Dimension Vista® CHEM 1 CAL
- Generic Name
- Multiple-type clinical chemistry analyte profile IVD, calibrator
- Model / Ref / Catalog
- 10716281, KC110B
- Primary DI / UDI-DI
- 00630414948508
- 510(k) Number
- FDA Product Code
- JIX
- GMDN Term
- Multiple-type clinical chemistry analyte profile IVD, calibrator
- Device Class
- Class II
- Regulation Number
- 862.1150
- GUDID distribution status
- Not in Commercial Distribution
Other devices from Siemens Healthcare Diagnostics Inc.
Properties
Other properties (12)
- Sterile: No
- Single Use: No
- Implantable: No
- Contains Latex: No
- Prescription Use: No
- Over-the-Counter: No
- Convenience Kit: No
- Combination Product: No
- HCT/P: No
- Premarket Exempt: No
- Serial Number: No
- Manufacturing Date: No