  "format": "devicerts.device.v1",
  "cite_as": "Discovery 670 DR, GE Medical Systems Israel, Functional Imaging. deviCERTS medical device catalogue. https://devicerts.com/devices/discovery-670-dr-0a6d3033 (data as of 2026-11-28).",
  "source": {
    "canonical_url": "https://devicerts.com/devices/discovery-670-dr-0a6d3033",
    "json_url": "https://devicerts.com/devices/discovery-670-dr-0a6d3033.json",
    "markdown_url": "https://devicerts.com/devices/discovery-670-dr-0a6d3033.md",
    "publisher": "deviCERTS (devicerts.com)",
    "data_sources": [
      {
        "registry": "FDA GUDID (AccessGUDID)",
        "url": "https://accessgudid.nlm.nih.gov/"
      },
      {
        "registry": "FDA recalls",
        "url": "https://open.fda.gov/apis/device/recall/"
      }
    ],
    "registry_records_last_synced": "2026-10-23",
    "status_as_of": "2026-11-28",
    "license": "CC BY 4.0",
    "license_url": "https://creativecommons.org/licenses/by/4.0/",
    "license_scope": "The deviCERTS layout and text are licensed under CC BY 4.0. Registry data remain under their source terms: U.S. FDA data (openFDA, AccessGUDID) are U.S. public domain; EUDAMED data are subject to the European Commission reuse terms.",
    "attribution": "DevicErts, https://devicerts.com/devices/discovery-670-dr-0a6d3033",
    "disclaimer": "Compiled by deviCERTS (devicerts.com) from public medical device registries (EUDAMED, European Commission; U.S. FDA via openFDA; AccessGUDID, U.S. National Library of Medicine). deviCERTS is not an official register and gives no regulatory advice: the record in the registry itself prevails."
  },
  "device": {
    "name": "Discovery 670 DR",
    "registry_name": "Discovery 670 DR",
    "url": "https://devicerts.com/devices/discovery-670-dr-0a6d3033",
    "manufacturer": {
      "name": "GE Medical Systems Israel, Functional Imaging",
      "url": "https://devicerts.com/manufacturers/ge-medical-systems-israel-functional-imaging",
      "country": "Israel",
      "country_basis": null,
      "fda_registration_country": null
    },
    "status": {
      "state": "active",
      "label": "Active",
      "detail": "United States"
    },
    "classification": [
      {
        "class": "Class II",
        "system": "US FDA"
      }
    ],
    "identifiers": [
      {
        "label": "Primary DI / UDI-DI",
        "values": [
          {
            "value": "00840682139359",
            "registries": [
              "FDA GUDID (AccessGUDID)"
            ]
          }
        ]
      },
      {
        "label": "510(k) Number",
        "values": [
          {
            "value": "K093514",
            "registries": [
              "FDA GUDID (AccessGUDID)"
            ]
          }
        ]
      }
    ],
    "product_codes": [
      "KPS",
      "JAK"
    ],
    "emdn": [],
    "gmdn": [
      {
        "code": "40640",
        "term": "Stationary gamma camera system"
      }
    ],
    "description": null,
    "noindex": true
  },
  "authorizations": [
    {
      "registry": "GUDID",
      "region": "United States",
      "type": "GUDID listing",
      "number": "00840682139359",
      "number_label": "UDI-DI",
      "device_udi_di": null,
      "date": "2018-05-15",
      "date_kind": "listed",
      "decision_code": null,
      "expires": null,
      "withdrawn": null,
      "status": "Listed",
      "status_note": "GUDID: in commercial distribution",
      "derived_from_gudid": false
    },
    {
      "registry": "GUDID",
      "region": "United States",
      "type": "FDA 510(k)",
      "number": "K093514",
      "number_label": null,
      "device_udi_di": null,
      "date": "2009-12-10",
      "date_kind": "decision",
      "decision_code": "SESE",
      "expires": null,
      "withdrawn": null,
      "status": "Active",
      "status_note": null,
      "derived_from_gudid": true
    }
  ],
  "no_longer_listed_by": [],
  "certificates": [],
  "recalls": [
    {
      "number": "Z-1053-2023",
      "classification": "Class I",
      "status": "ongoing",
      "initiated": "2022-12-19",
      "reason": "A mitigation may not be correctly implemented. If this is the case, the detector can fall, leading to life-threatening bodily injury.",
      "applies": "this device",
      "via": "K093514"
    },
    {
      "number": "Z-0891-2025",
      "classification": "Class II",
      "status": "ongoing",
      "initiated": "2024-12-18",
      "reason": "GE HealthCare has become aware that certain Infinia, VariCam/VG/VH, Brivo NM 615, Discovery NM 630, Discovery NM/CT 670, and Optima NM/CT 640 Nuclear Medicine systems (see Affected Products Details and Appendices) could have been transported or relocated without proper detector support, which can result in excessive load being applied to the detector mounting mechanisms. If this occurs, it can potentially result in a detector fall and lead to life-threatening bodily injury.",
      "applies": "this device",
      "via": "K093514"
    },
    {
      "number": "Z-0893-2025",
      "classification": "Class II",
      "status": "ongoing",
      "initiated": "2024-12-18",
      "reason": "GE HealthCare has become aware that certain Infinia, VariCam/VG/VH, Brivo NM 615, Discovery NM 630, Discovery NM/CT 670, and Optima NM/CT 640 Nuclear Medicine systems (see Affected Products Details and Appendices) could have been transported or relocated without proper detector support, which can result in excessive load being applied to the detector mounting mechanisms. If this occurs, it can potentially result in a detector fall and lead to life-threatening bodily injury.",
      "applies": "this device",
      "via": "K093514"
    },
    {
      "number": "Z-0894-2025",
      "classification": "Class II",
      "status": "ongoing",
      "initiated": "2024-12-18",
      "reason": "GE HealthCare has become aware that certain Infinia, VariCam/VG/VH, Brivo NM 615, Discovery NM 630, Discovery NM/CT 670, and Optima NM/CT 640 Nuclear Medicine systems (see Affected Products Details and Appendices) could have been transported or relocated without proper detector support, which can result in excessive load being applied to the detector mounting mechanisms. If this occurs, it can potentially result in a detector fall and lead to life-threatening bodily injury.",
      "applies": "this device",
      "via": "K093514"
    },
    {
      "number": "Z-0895-2025",
      "classification": "Class II",
      "status": "ongoing",
      "initiated": "2024-12-18",
      "reason": "GE HealthCare has become aware that certain Infinia, VariCam/VG/VH, Brivo NM 615, Discovery NM 630, Discovery NM/CT 670, and Optima NM/CT 640 Nuclear Medicine systems (see Affected Products Details and Appendices) could have been transported or relocated without proper detector support, which can result in excessive load being applied to the detector mounting mechanisms. If this occurs, it can potentially result in a detector fall and lead to life-threatening bodily injury.",
      "applies": "this device",
      "via": "K093514"
    },
    {
      "number": "Z-1547-2021",
      "classification": "Class II",
      "status": "ongoing",
      "initiated": "2021-02-07",
      "reason": "GE Healthcare has become aware of a potential issue on 800 family NM/CT\nScanners that could cause a shift in CT radiation exposure range of up to 5cm compared to the\nintended radiation exposure range of the planned scan under specific workflows. The issue occurs\nonly on a hybrid whole-body continuous F3 protocol with Zoom < 1 where the scan range is set on\nthe Smart Console. In some cases, this may also necessitate a re-scan of the patient, which would\nexpose the patient to additional X-ray radiation.",
      "applies": "this device",
      "via": "K093514"
    },
    {
      "number": "Z-0857-2021",
      "classification": "Class II",
      "status": "terminated",
      "initiated": "2020-12-10",
      "reason": "GE Healthcare has become aware that rotor bearing screws were found loose on one detector in the field, leading to the release of one of the four rotor bearings in the detector.",
      "applies": "this device",
      "via": "K093514"
    },
    {
      "number": "Z-2996-2020",
      "classification": "Class II",
      "status": "terminated",
      "initiated": "2020-08-14",
      "reason": "Rotor bearing screws were found loose on\ndetector.",
      "applies": "this device",
      "via": "K093514"
    },
    {
      "number": "Z-2997-2020",
      "classification": "Class II",
      "status": "terminated",
      "initiated": "2020-08-14",
      "reason": "Rotor bearing screws were found loose on\ndetector.",
      "applies": "this device",
      "via": "K093514"
    },
    {
      "number": "Z-2998-2020",
      "classification": "Class II",
      "status": "terminated",
      "initiated": "2020-08-14",
      "reason": "Rotor bearing screws were found loose on\ndetector.",
      "applies": "this device",
      "via": "K093514"
    },
    {
      "number": "Z-1259-2019",
      "classification": "Class II",
      "status": "terminated",
      "initiated": "2019-03-20",
      "reason": "During installation of certain Nuclear Medicine systems delivered with dismounted detectors, some detector screws may not have been secured correctly, leading to the possible loss of one detector mounting design redundancy.",
      "applies": "this device",
      "via": "K093514"
    },
    {
      "number": "Z-1260-2019",
      "classification": "Class II",
      "status": "terminated",
      "initiated": "2019-03-20",
      "reason": "During installation of certain Nuclear Medicine systems delivered with dismounted detectors, some detector screws may not have been secured correctly, leading to the possible loss of one detector mounting design redundancy.",
      "applies": "this device",
      "via": "K093514"
    },
    {
      "number": "Z-1265-2019",
      "classification": "Class II",
      "status": "terminated",
      "initiated": "2019-03-20",
      "reason": "During installation of certain Nuclear Medicine systems delivered with dismounted detectors, some detector screws may not have been secured correctly, leading to the possible loss of one detector mounting design redundancy.",
      "applies": "this device",
      "via": "K093514"
    },
    {
      "number": "Z-0606-2019",
      "classification": "Class II",
      "status": "completed",
      "initiated": "2018-05-24",
      "reason": "When your system was installed, the Main disconnect switch, EPO button on the A1 panel, and Remote EPO button(s) (outside of the A1 panel) may not have been wired correctly which could result in these not shutting off all power to the entire system as intended, resulting in a potential electric hazard.",
      "applies": "this device",
      "via": "K093514"
    },
    {
      "number": "Z-0607-2019",
      "classification": "Class II",
      "status": "completed",
      "initiated": "2018-05-24",
      "reason": "When your system was installed, the Main disconnect switch, EPO button on the A1 panel, and Remote EPO button(s) (outside of the A1 panel) may not have been wired correctly which could result in these not shutting off all power to the entire system as intended, resulting in a potential electric hazard.",
      "applies": "this device",
      "via": "K093514"
    },
    {
      "number": "Z-0608-2019",
      "classification": "Class II",
      "status": "completed",
      "initiated": "2018-05-24",
      "reason": "When your system was installed, the Main disconnect switch, EPO button on the A1 panel, and Remote EPO button(s) (outside of the A1 panel) may not have been wired correctly which could result in these not shutting off all power to the entire system as intended, resulting in a potential electric hazard.",
      "applies": "this device",
      "via": "K093514"
    },
    {
      "number": "Z-2194-2017",
      "classification": "Class II",
      "status": "terminated",
      "initiated": "2017-05-19",
      "reason": "It was discovered that an issue has been identified that can result in a scan starting before the prescribed delay has completed.",
      "applies": "this device",
      "via": "K093514"
    },
    {
      "number": "Z-0367-2014",
      "classification": "Class II",
      "status": "terminated",
      "initiated": "2013-10-25",
      "reason": "GE Healthcare has recently become aware of a potential safety issue related to a SPECT scan procedure used for pulmonary imaging.\n\nContact between the Detector (GE Healthcare Discovery NM/CT 670, and/or Optima NM/CT 640, and/or Discovery\nNM630, and/or Brivo NM 615 all configurations) and the Patient's elbow may occur during SPECT protocols using 450 angular steps in scenarios when a Patients a",
      "applies": "this device",
      "via": "K093514"
    },
    {
      "number": "Z-1792-2013",
      "classification": "Class I",
      "status": "terminated",
      "initiated": "2013-06-13",
      "reason": "GE became aware of an incident at a VA Medical Center facility in the US in which a patient died due to injuries sustained while being scanned on an Infinia Hawkeye 4 Nuclear Medicine System.  UPDATE:  On July 03, 2013 GE notified hospitals that they are recalling several Nuclear Medicine Imaging Systems because serious injuries or deaths could occur due to the failure mode associated with this re",
      "applies": "this device",
      "via": "K093514"
    },
    {
      "number": "Z-0121-2013",
      "classification": "Class II",
      "status": "terminated",
      "initiated": "2012-10-12",
      "reason": "GE Healthcare has recently become aware of a potential safety issue associated with the\ntable pallet of your Discovery NM/CT 670 and/or Discovery NM630 system which may impact\noperation safety.\n\nAt scan setup, during gantry rotation towards scanning position, a potential contact between the table-pallet-edge and the detector was identified. Such occurrence may damage the pallet front-edge and",
      "applies": "this device",
      "via": "K093514"
    },
    {
      "number": "Z-1110-2012",
      "classification": "",
      "status": "terminated",
      "initiated": "2012-01-11",
      "reason": "GE Healthcare has recently become aware of a potential safety issue associated with the collimator cart and lead cover of your Discovery NM/CT 670 and/or Discovery NM 630 system that may impact operator safety. \n\n1.\tThe cart locking pin release and cart disengagement could result in the collimator falling down and could lead to operator or service personnel injury.\n2.\tAnother issue is that the",
      "applies": "this device",
      "via": "K093514"
    }
  ],
  "eu_market_availability": null,
  "properties": [
    {
      "group": "Identity",
      "fields": [
        {
          "label": "Trade name",
          "values": [
            {
              "value": "Discovery",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        },
        {
          "label": "Generic name",
          "values": [
            {
              "value": "Stationary gamma camera system",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        },
        {
          "label": "Model / reference",
          "values": [
            {
              "value": "670 DR",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        }
      ]
    },
    {
      "group": "Identifiers",
      "fields": [
        {
          "label": "Primary DI / UDI-DI",
          "values": [
            {
              "value": "00840682139359",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        },
        {
          "label": "510(k) Number",
          "values": [
            {
              "value": "K093514",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        },
        {
          "label": "FDA product code",
          "values": [
            {
              "value": "KPS",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            },
            {
              "value": "JAK",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        },
        {
          "label": "GMDN term",
          "values": [
            {
              "value": "Stationary gamma camera system",
              "code": "40640",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        }
      ]
    },
    {
      "group": "Classification",
      "fields": [
        {
          "label": "Device class",
          "values": [
            {
              "value": "Class II",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        },
        {
          "label": "Regulation number",
          "values": [
            {
              "value": "892.1200",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            },
            {
              "value": "892.1750",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        },
        {
          "label": "GUDID distribution status",
          "values": [
            {
              "value": "In Commercial Distribution",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        }
      ]
    },
    {
      "group": "Properties",
      "fields": [
        {
          "label": "Prescription use",
          "values": [
            {
              "value": "Yes",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        },
        {
          "label": "Serial number",
          "values": [
            {
              "value": "Yes",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        },
        {
          "label": "Manufacturing date",
          "values": [
            {
              "value": "Yes",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        },
        {
          "label": "Supplied sterile",
          "values": [
            {
              "value": "No",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        },
        {
          "label": "Sterilise before use",
          "values": [
            {
              "value": "No",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        },
        {
          "label": "Single use",
          "values": [
            {
              "value": "No",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        },
        {
          "label": "Implantable",
          "values": [
            {
              "value": "No",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        },
        {
          "label": "Contains latex",
          "values": [
            {
              "value": "No",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        },
        {
          "label": "Over-the-counter",
          "values": [
            {
              "value": "No",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        },
        {
          "label": "Convenience kit",
          "values": [
            {
              "value": "No",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        },
        {
          "label": "Combination product",
          "values": [
            {
              "value": "No",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        },
        {
          "label": "HCT/P",
          "values": [
            {
              "value": "No",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        },
        {
          "label": "Premarket exempt",
          "values": [
            {
              "value": "No",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        },
        {
          "label": "Lot / batch number",
          "values": [
            {
              "value": "No",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        },
        {
          "label": "Expiration date",
          "values": [
            {
              "value": "No",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        }
      ]
    }
  ]
}