  "format": "devicerts.device.v1",
  "cite_as": "Discovery Nm 630, GE Healthcare. deviCERTS medical device catalogue. https://devicerts.com/devices/discovery-nm-630-kps (data as of 2026-11-20).",
  "source": {
    "canonical_url": "https://devicerts.com/devices/discovery-nm-630-kps",
    "json_url": "https://devicerts.com/devices/discovery-nm-630-kps.json",
    "markdown_url": "https://devicerts.com/devices/discovery-nm-630-kps.md",
    "publisher": "deviCERTS (devicerts.com)",
    "data_sources": [
      {
        "registry": "FDA 510(k)",
        "url": "https://open.fda.gov/apis/device/510k/"
      },
      {
        "registry": "FDA recalls",
        "url": "https://open.fda.gov/apis/device/recall/"
      }
    ],
    "registry_records_last_synced": "2026-10-22",
    "status_as_of": "2026-11-20",
    "license": "CC BY 4.0",
    "license_url": "https://creativecommons.org/licenses/by/4.0/",
    "license_scope": "The deviCERTS layout and text are licensed under CC BY 4.0. Registry data remain under their source terms: U.S. FDA data (openFDA, AccessGUDID) are U.S. public domain; EUDAMED data are subject to the European Commission reuse terms.",
    "attribution": "DevicErts, https://devicerts.com/devices/discovery-nm-630-kps",
    "disclaimer": "Compiled by deviCERTS (devicerts.com) from public medical device registries (EUDAMED, European Commission; U.S. FDA via openFDA; AccessGUDID, U.S. National Library of Medicine). deviCERTS is not an official register and gives no regulatory advice: the record in the registry itself prevails."
  },
  "device": {
    "name": "Discovery Nm 630",
    "registry_name": "DISCOVERY NM 630",
    "url": "https://devicerts.com/devices/discovery-nm-630-kps",
    "manufacturer": {
      "name": "GE Healthcare",
      "url": "https://devicerts.com/manufacturers/ge-healthcare",
      "country": "United States",
      "country_basis": null,
      "fda_registration_country": null
    },
    "status": {
      "state": "active",
      "label": "Active",
      "detail": "United States"
    },
    "classification": [
      {
        "class": "Class II",
        "system": "US FDA"
      }
    ],
    "identifiers": [
      {
        "label": "510(k) Number",
        "values": [
          {
            "value": "K111445",
            "registries": [
              "FDA 510(k)"
            ]
          }
        ]
      }
    ],
    "product_codes": [
      "KPS"
    ],
    "emdn": [],
    "gmdn": [],
    "description": null,
    "noindex": false
  },
  "authorizations": [
    {
      "registry": "510(k)",
      "region": "United States",
      "type": "FDA 510(k)",
      "number": "K111445",
      "number_label": null,
      "device_udi_di": null,
      "date": "2011-06-03",
      "date_kind": "decision",
      "decision_code": "SESE",
      "expires": null,
      "withdrawn": null,
      "status": "Active",
      "status_note": null,
      "derived_from_gudid": false
    }
  ],
  "no_longer_listed_by": [],
  "certificates": [],
  "recalls": [
    {
      "number": "Z-0890-2025",
      "classification": "Class II",
      "status": "ongoing",
      "initiated": "2024-12-18",
      "reason": "GE HealthCare has become aware that certain Infinia, VariCam/VG/VH, Brivo NM 615, Discovery NM 630, Discovery NM/CT 670, and Optima NM/CT 640 Nuclear Medicine systems (see Affected Products Details and Appendices) could have been transported or relocated without proper detector support, which can result in excessive load being applied to the detector mounting mechanisms. If this occurs, it can potentially result in a detector fall and lead to life-threatening bodily injury.",
      "applies": "this device",
      "via": "K111445"
    },
    {
      "number": "Z-1054-2023",
      "classification": "Class I",
      "status": "ongoing",
      "initiated": "2022-12-19",
      "reason": "A mitigation may not be correctly implemented. If this is the case, the detector can fall, leading to life-threatening bodily injury.",
      "applies": "this device",
      "via": "K111445"
    },
    {
      "number": "Z-0856-2021",
      "classification": "Class II",
      "status": "terminated",
      "initiated": "2020-12-10",
      "reason": "GE Healthcare has become aware that rotor bearing screws were found loose on one detector in the field, leading to the release of one of the four rotor bearings in the detector.",
      "applies": "this device",
      "via": "K111445"
    },
    {
      "number": "Z-2994-2020",
      "classification": "Class II",
      "status": "terminated",
      "initiated": "2020-08-14",
      "reason": "Rotor bearing screws were found loose on\ndetector.",
      "applies": "this device",
      "via": "K111445"
    },
    {
      "number": "Z-2995-2020",
      "classification": "Class II",
      "status": "terminated",
      "initiated": "2020-08-14",
      "reason": "Rotor bearing screws were found loose on\ndetector.",
      "applies": "this device",
      "via": "K111445"
    },
    {
      "number": "Z-1257-2019",
      "classification": "Class II",
      "status": "terminated",
      "initiated": "2019-03-20",
      "reason": "During installation of certain Nuclear Medicine systems delivered with dismounted detectors, some detector screws may not have been secured correctly, leading to the possible loss of one detector mounting design redundancy.",
      "applies": "shared submission",
      "via": "K111445"
    },
    {
      "number": "Z-0367-2014",
      "classification": "Class II",
      "status": "terminated",
      "initiated": "2013-10-25",
      "reason": "GE Healthcare has recently become aware of a potential safety issue related to a SPECT scan procedure used for pulmonary imaging.\n\nContact between the Detector (GE Healthcare Discovery NM/CT 670, and/or Optima NM/CT 640, and/or Discovery\nNM630, and/or Brivo NM 615 all configurations) and the Patient's elbow may occur during SPECT protocols using 450 angular steps in scenarios when a Patients a",
      "applies": "this device",
      "via": "K111445"
    },
    {
      "number": "Z-1791-2013",
      "classification": "Class I",
      "status": "terminated",
      "initiated": "2013-06-13",
      "reason": "GE became aware of an incident at a VA Medical Center facility in the US in which a patient died due to injuries sustained while being scanned on an Infinia Hawkeye 4 Nuclear Medicine System.  UPDATE:  On July 03, 2013 GE notified hospitals that they are recalling several Nuclear Medicine Imaging Systems because serious injuries or deaths could occur due to the failure mode associated with this re",
      "applies": "this device",
      "via": "K111445"
    },
    {
      "number": "Z-0121-2013",
      "classification": "Class II",
      "status": "terminated",
      "initiated": "2012-10-12",
      "reason": "GE Healthcare has recently become aware of a potential safety issue associated with the\ntable pallet of your Discovery NM/CT 670 and/or Discovery NM630 system which may impact\noperation safety.\n\nAt scan setup, during gantry rotation towards scanning position, a potential contact between the table-pallet-edge and the detector was identified. Such occurrence may damage the pallet front-edge and",
      "applies": "this device",
      "via": "K111445"
    },
    {
      "number": "Z-1111-2012",
      "classification": "",
      "status": "terminated",
      "initiated": "2012-01-11",
      "reason": "GE Healthcare has recently become aware of a potential safety issue associated with the collimator cart and lead cover of your Discovery NM/CT 670 and/or Discovery NM 630 system that may impact operator safety. \n\n1.\tThe cart locking pin release and cart disengagement could result in the collimator falling down and could lead to operator or service personnel injury.\n2.\tAnother issue is that the",
      "applies": "this device",
      "via": "K111445"
    }
  ],
  "eu_market_availability": null,
  "properties": [
    {
      "group": "Identifiers",
      "fields": [
        {
          "label": "510(k) Number",
          "values": [
            {
              "value": "K111445",
              "registries": [
                "FDA 510(k)"
              ]
            }
          ]
        },
        {
          "label": "FDA product code",
          "values": [
            {
              "value": "KPS",
              "registries": [
                "FDA 510(k)"
              ]
            }
          ]
        }
      ]
    },
    {
      "group": "Classification",
      "fields": [
        {
          "label": "Device class",
          "values": [
            {
              "value": "Class II",
              "registries": [
                "FDA 510(k)"
              ]
            }
          ]
        },
        {
          "label": "Regulation",
          "values": [
            {
              "value": "21 CFR Part 892.1200",
              "registries": [
                "FDA 510(k)"
              ]
            }
          ]
        },
        {
          "label": "Advisory committee",
          "values": [
            {
              "value": "Radiology",
              "registries": [
                "FDA 510(k)"
              ]
            }
          ]
        },
        {
          "label": "FDA decision",
          "values": [
            {
              "value": "SESE",
              "registries": [
                "FDA 510(k)"
              ]
            }
          ]
        }
      ]
    }
  ]
}