Djo Surgical 520-54-320
Djo Surgical 520-54-320 is a device of GMDN type total shoulder prosthesis (FDA Class II) made by Encore Medical LP, cleared under FDA 510(k) K162024 on 21 Nov 2016 and listed in FDA GUDID since 17 Jan 2017 (UDI-DI 00190446176275), active as of 8 Nov 2026. 6 FDA recalls are linked to it.
Active
Valid until
No expiry
FDA recalls
1 ongoingDocuments
Trust
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | GUDID listing | Number00190446176275 | Date17 Jan 2017 | StatusListed |
| United States | FDA 510(k)SESE | NumberK162024 | Date21 Nov 2016 | StatusActive |
Official records
- AccessGUDID record 00190446176275
- FDA 510(k) database K162024
From the FDA record
- Device description
- AltiVate Anatomic, Offset Humeral Head, 54X20
- MRI safety
- Labeling does not contain MRI Safety Information
- Natural rubber latex
- Labelled as not made with natural rubber latex
- Sterility
- Supplied sterile
Across registries
- FDA 510(k) K162024 decided 21 Nov 2016 Active
- FDA GUDID UDI-DI 00190446176275 listed 17 Jan 2017 Listed
Recalls
-
- Ongoing Z-2331-2021
Complaints have been received concerning the humeral stem failing to mate with the broach handles for a shoulder device. This could result in delay of procedure or revision surgery.
-
- Completed Z-1256-2018
- Completed Z-1257-2018
- Completed Z-1258-2018
- Completed Z-1259-2018
- Completed Z-1260-2018
It was discovered during a surgery that the design of the radius at the base of the pegs of the Anatomic Glenoid Trial (Part Numbers: 804-07-380, 804-07-420, 804-07-460, 804-07-500, and 804-07-540) is much larger on the trails than on the implants
Timeline
- Cleared FDA 510(k) K162024
- Listed FDA GUDID 00190446176275
- Recall initiated FDA recalls Z-1256-2018
- Recall initiated FDA recalls Z-1257-2018
- Recall initiated FDA recalls Z-1258-2018
- Recall initiated FDA recalls Z-1259-2018
- Recall initiated FDA recalls Z-1260-2018
- Recall initiated FDA recalls Z-2331-2021
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- Trade name
- Djo Surgical
- Generic name
- Total shoulder prosthesis
- Model / reference
- 520-54-320
- Primary DI / UDI-DI
- 00190446176275
- 510(k) Number
- FDA product code
- PAO, HSD, KWS
- GMDN term
- Total shoulder prosthesis
- Device class
- Class II
- Regulation number
- 888.3660, 888.3690
- GUDID distribution status
- In Commercial Distribution
Similar devices in this category