Djo Surgical 804-07-002
Djo Surgical 804-07-002 is a medical device (FDA Class II) made by Encore Medical LP, cleared under FDA 510(k) K162024 on 21 Nov 2016 and listed in FDA GUDID since 17 Jan 2017 (UDI-DI 00190446173953), active as of 8 Nov 2026. 6 FDA recalls are linked to it.
Active
Valid until
No expiry
FDA recalls
1 ongoingDocuments
Trust
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | GUDID listing | Number00190446173953 | Date17 Jan 2017 | StatusListed |
| United States | FDA 510(k)SESE | NumberK162024 | Date21 Nov 2016 | StatusActive |
Official records
- AccessGUDID record 00190446173953
- FDA 510(k) database K162024
From the FDA record
- Device description
- AltiVate Anatomic, Humeral Head Neck Adaptor, Offset
- MRI safety
- Labeling does not contain MRI Safety Information
- Natural rubber latex
- Labelled as not made with natural rubber latex
- Sterility
- Supplied non-sterile; sterilise before use (Moist Heat or Steam Sterilization)
Across registries
- FDA 510(k) K162024 decided 21 Nov 2016 Active
- FDA GUDID UDI-DI 00190446173953 listed 17 Jan 2017 Listed
Recalls
-
- Ongoing Z-2331-2021
Complaints have been received concerning the humeral stem failing to mate with the broach handles for a shoulder device. This could result in delay of procedure or revision surgery.
-
- Completed Z-1256-2018
- Completed Z-1257-2018
- Completed Z-1258-2018
- Completed Z-1259-2018
- Completed Z-1260-2018
It was discovered during a surgery that the design of the radius at the base of the pegs of the Anatomic Glenoid Trial (Part Numbers: 804-07-380, 804-07-420, 804-07-460, 804-07-500, and 804-07-540) is much larger on the trails than on the implants
Timeline
- Cleared FDA 510(k) K162024
- Listed FDA GUDID 00190446173953
- Recall initiated FDA recalls Z-1256-2018
- Recall initiated FDA recalls Z-1257-2018
- Recall initiated FDA recalls Z-1258-2018
- Recall initiated FDA recalls Z-1259-2018
- Recall initiated FDA recalls Z-1260-2018
- Recall initiated FDA recalls Z-2331-2021
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- Trade name
- Djo Surgical
- Generic name
- Orthopaedic prosthesis implantation positioning instrument, reusable
- Model / reference
- 804-07-002
- Primary DI / UDI-DI
- 00190446173953
- 510(k) Number
- FDA product code
- PAO, HSD, KWS
- GMDN term
- Orthopaedic prosthesis implantation positioning instrument, reusable
- Device class
- Class II
- Regulation number
- 888.3660, 888.3690
- GUDID distribution status
- In Commercial Distribution
- Sterilisation method
- Moist Heat or Steam Sterilization
Similar devices in this category
Other devices filed under Orthopaedic prosthesis implantation positioning instrument, reusable.
- Vanguard® Microplasty® 32-485053 by Biomet Orthopedics, Inc.
- 00-5851-020-00 by Zimmer, Inc.
- 2410-20 00-2410-020-00 by Zimmer, Inc.
- AESCULAP NP605R by Aesculap AG
- ATTUNE 2546-00-426 254600426 by DEPUY (IRELAND)
- INSTRUMENT 6210-9-100 by Howmedica Osteonics Corp.
- VARIAX 703818 by Stryker GmbH
- MIOS NQ976R by Aesculap AG
- HYPE stem TE607- 32 CL by SOC ETUDES RECHERCHES FABRICATION
- Vanguard® 32-486000 by Biomet Orthopedics, Inc.