DL900 Series Holter Recorder
by Braemar, Inc. 510(k)
Active
510(k) K071733
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK071733 | Date24 Jul 2007 | StatusActive |
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not known
- Not covered
United States — FDA 510(k) Active
Recalls
-
- Terminated Z-1460-2020
- Terminated Z-1459-2020
An issue with the DL900 and DL950 Series Holter Recorder was recognized on January 1, 2020. This issue can prevent the recorder from operating normally. Braemar confirmed that beginning on January 1, 2020 if a AAA battery is inserted in the recorder and a user attempts to start it, the recorder will display Error: 602 and fail to function for a new patient study. There are no actions that a clinical user can take to clear this error. Even if this error code is displayed, existing ECG recordings stored on the device can still be downloaded via the standard Holter Application Software. Error 602 may be able to be cleared by technical or engineering staff at the customer site.
-
- Terminated Z-1345-2020
if an AAA battery is inserted in the recorder and a user attempts to start it, or if the recorder is inserted in the docking station, the recorder will display Error: 602 and fail to function for a new patient study.
Specification
- 510(k) Number
- FDA Product Code
- MWJ
- Device Class
- Class II
- Regulation
- 21 CFR Part 870.2800
- Advisory Committee
- Cardiovascular
- FDA Decision
- SESE Substantially Equivalent