Drake Willock System 1000 Dialysate Delivery Syst
510(k) K910215
Active
Valid until
No expiry
FDA recalls
None ongoingDocuments
Trust
Unverified
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK910215 | Date9 Apr 1991 | StatusActive |
Recalls
-
- Terminated Z-1253-06
The UF Removal Regulators were assembled incorrectly. The mis-assembly results in a failure to successfully calibrate the instrument and begin patient therapy.
-
- Terminated Z-0766-06
The older style 'Through Hardware' Dialysate Monitoring Manifolds and 'Noryl' Ultra Filtration Removal Regulators may become cracked due to mechanical and thermal stress. If either of these parts become cracked during patient treatment, there is the potential for an ultrafiltration error to occur, which can result in excessive removal of fluid from the patient.
-
- Terminated Z-0347-06
Potential for fluid or air to be passed through the venous line clamp if the tubing is not centered in the clamp and extends beyond the edge of the clamp's pinch zone. This could result in an air emboli condition.
-
- Terminated Z-0395-05
The air detector may not detect air bubbles consistently at the selected limit.
-
- Terminated Z-0833-03
The dialysis treatment may be run without the bicarbonate due to an undetected bicarbonate pump stall following a sodium change.
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- 510(k) Number
- FDA Product Code
- FKP
- Device Class
- Class II
- Regulation
- 21 CFR Part 876.5820
- Advisory Committee
- Gastroenterology, Urology
- FDA Decision
- SESE Substantially Equivalent