Recalls
-
- Terminated Z-2651-2017
Carestream found a single unit where four screws in the assembly were not applied in the manufacturing process. This could result in separation of the tube and yoke assembly.
-
- Terminated Z-1138-2017
The firm received a complaint alleging that a DRX Revolution System could not be stopped when applying the brake. The Revolution crashed into an elevator.
-
- Terminated Z-0800-2015
While servicing a unit at a customer site, a field engineer discovered a broken bolt from the carriage assembly. Carestream has identified an issue related to the DRX-Revolution Mobile X-Ray System in which bolts that secure the x-ray tube assembly/boom to the device mounting block may break. It has been determined that this issue is a result of an improperly assembled part which holds the x-ray
-
- Terminated Z-0179-2014
Carestream Health Inc. has conducted a voluntary field action which includes the installation of a software patch on the following: DRX-Revolution Mobile X-Ray, DRX-Mobile and DRX Transportable Systems with DirectView Version 5.5 and 5.6 console software when used with DRX-1 Detector version 69 firmware. Carestream recently received reports of a non-diagnostic image on exams when a Normal Exposure
-
- Terminated Z-0090-2014
During an audit, Carestream discovered a DRX-1 System that captured an image which displayed an artifact. This issue was discovered in manufacturing. It occurs when the DRX product console software is updated to v5.6b software and the detector is not calibrated before use.
-
- Terminated Z-0132-2014
Carestream Health Inc. is conducting a recall for the Column end Cover present on the DRX Revolution Mobile X-Ray System due to the cover dislodging from the column.
-
- Terminated Z-0126-2014
Carestream Health Inc. is recalling the power plug on their DRX-Revolution Mobile X-Ray System due to the item generating sparks and becoming warm to touch.
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — GUDID listing Listed
Specification
- Generic Name
- Mobile basic diagnostic x-ray system, digital
- Model / Ref / Catalog
- DRX-Revolution Mobile X-ray System, 8007254
- Primary DI / UDI-DI
- 60889978007257
- FDA Product Code
- IZL
- GMDN Term
- Mobile basic diagnostic x-ray system, digital
- Device Class
- Class II
- Regulation Number
- 892.1720
- GUDID distribution status
- Not in Commercial Distribution
Other devices from Carestream Health, Inc.
Properties
Other properties (12)
- Sterile: No
- Single Use: No
- Implantable: No
- Contains Latex: No
- Over-the-Counter: No
- Convenience Kit: No
- Combination Product: No
- HCT/P: No
- Premarket Exempt: No
- Lot / Batch Number: No
- Expiration Date: No
- Manufacturing Date: No