Dsl Active Igfbp-3 Irma (Dsl 6600)

by Diagnostic Systems Laboratories, Inc. 510(k)

Active United States

510(k) K950228

Regulatory status

as of 28 Oct 2026

RegionAuthorizationNumberDateStatus
United StatesFDA 510(k)SESE Substantially Equivalent · Traditional NumberK950228 Date22 Feb 1995ClearedStatusActive

Coverage

1 jurisdiction

United States of America
  • Active
  • Expiring
  • Withdrawn
  • Listed
  • Not covered

United States — FDA 510(k) Active

Recalls

1 recall · FDA

    • Terminated Z-2607-2010 FDA Unclassified (pre-2012) initiated 9 Aug 2010

    The recall was initiated after Beckman Coulter confirmed customer reports that IGFBP-3 measurements using IGFBP-3 IRMA kit (REF DSL-6600) lots identified above may result in IGFBP-3 levels that are not in agreement with simultaneously measured levels of IGF-1. The investigation confirmed that above mentioned lots of IGFBP-3 kit provide results elevated by 84% on average. Erroneously elevated IGF

Specification

510(k)

510(k) Number
FDA Product Code
CFL
Device Class
Class I
Regulation
21 CFR Part 862.1370
Advisory Committee
Clinical Chemistry
FDA Decision
SESE Substantially Equivalent