Dsl Estrone (Dsl 8700)
by Diagnostic Systems Laboratories, Inc. 510(k)
Active
510(k) K935013
Regulatory status
Recall ongoing. 3 ongoing FDA recalls. See recalls
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK935013 | Date16 Jun 1994 | StatusActive |
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not known
- Not covered
United States — FDA 510(k) Active
Recalls
-
- Ongoing Z-1369-2026
The affected Estrone RIA kit lots may rarely give falsely high patient results which might cause Temporary injury of Remote probability. Only in some (very rare) cases may the issue occur.
-
- Ongoing Z-2336-2025
Estrone RIA kit lots may rarely give falsely high patient results which might cause Temporary injury of Remote probability or rarely give falsely low patient results which might cause Permanent injury of Remote probability. Only in some (very rare) cases may the issue occur.
-
- Ongoing Z-2152-2025
A specific lot of bovine serum albumin (BSA) used for manufacturing of coated tubes (CT) that are included in the Estrone RIA kit lots caused different affinity of manufactured coated tubes to patient samples, causing falsely increased assay results for the impacted Estrone RIA kits.
-
- Terminated Z-1685-2011
Beckman Coulter has confirmed that eight Estrone RIA kit (Ref DSL 8700) lots are inaccurately calibrated to the firm's internal reference calibrator. Lots 992637, 100115A, 100205A, 100205RA, and 100226A have now expired and should not be in active use any more.
Specification
- 510(k) Number
- FDA Product Code
- CGF
- Device Class
- Class I
- Regulation
- 21 CFR Part 862.1280
- Advisory Committee
- Clinical Chemistry
- FDA Decision
- SESE Substantially Equivalent
Other devices from Diagnostic Systems Laboratories, Inc.
See all devices from Diagnostic Systems Laboratories, Inc. →