Duo Flow Catheter-Mc 54
by Medical Components, Inc. United States 510(k)
510(k) K842416 Product code KNZ
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK842416 | Date26 Jul 1984 | StatusActive |
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not known
- Not covered
United States — FDA 510(k) Active
Recalls
-
- Terminated Z-2942-2018
- Terminated Z-2937-2018
- Terminated Z-2935-2018
- Terminated Z-2941-2018
- Terminated Z-2940-2018
- Terminated Z-2936-2018
- Terminated Z-2939-2018
- Terminated Z-2933-2018
- Terminated Z-2934-2018
- Terminated Z-2938-2018
- Terminated Z-2932-2018
The suture wings have the potential to crack or break. Potential patient injury includes catheter dislocation, bleeding, and air emboli. The firm has received six complaints (29 broken suture wings) to date for malfunction, without serious injury or death.
Specification
- 510(k) Number
- FDA Product Code
- KNZ
- Device Class
- Class II
- Regulation
- 21 CFR Part 876.5540
- Advisory Committee
- Gastroenterology, Urology
- FDA Decision
- SESE Substantially Equivalent
From the manufacturer
Manufacturer's product page: Duo-Flow®
Other devices from Medical Components, Inc.
Properties & raw data
Show properties, registry sources and raw records
Sources (1)
- FDA 510(k) K842416