DxH 900 Hematology System C11478
Ref C11478
Active
Valid until
Not listed
FDA recalls
2 ongoingDocuments
None linked in EUDAMED
Trust
Unverified
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| European Union | CE/IVDD EUDAMED | NumberB-C11478UC D-C11478UC | Date8 Jun 2022 | StatusActive |
| United States | GUDID listing GUDID | NumberD-C11478UC | Date3 May 2018 | StatusListed |
Certificates
EUDAMED lists no certificate for this device's Basic UDI-DI.
Recalls
-
- Ongoing Z-2573-2025
- Ongoing Z-2572-2025
Hematology analyzers configured with HGB photometers may result in erroneously high hemoglobin (HGB) results on samples with elevated white blood cell (WBC) counts.
-
- Terminated Z-0516-2020
When the UniCel DxH Slidemaker Stainer and UniCel DxH Slidemaker Stainer II dispense probe is slightly bent, there is a potential for cell carryover from the previous sample, and erroneous results could be reported from the slide review. Erroneous results may be reported from the laboratory which may trigger unnecessary or inappropriate patient treatment.
-
- Terminated Z-1383-2019
- Terminated Z-1384-2019
- Terminated Z-1382-2019
Complaints received of sporadic erroneously elevated platelet results without flags or system messages. Thrombocytopenia may go unrecognized.
-
- Terminated Z-0236-2018
- Terminated Z-0238-2018
- Terminated Z-0237-2018
The shield for the Single-Tube Station on the DxH instrument may interfere with sample aspiration in the open vial position.
-
- Terminated Z-3101-2017
The firm has identified that a fitting in some Slidemaker Stainer units may not completely seal after installation.
-
- Terminated Z-2158-2017
- Terminated Z-2157-2017
There is a possibility of a data acquisition disruption that may cause an unusual light scatter event pattern for the white blood cell differential that may not have a system flag and/or message. The events associated with the unusual light scatter pattern may be incorrectly removed from analysis which can result in an erroneous differential result.
-
- Terminated Z-1013-2017
- Terminated Z-1014-2017
Beckman Coulter has determined that additional clarification for the Blast Suspect messages is necessary. In rare situations, the UniCel DxH 800 and DxH 600 Coulter Cellular Analysis System may not flag or detect blasts in some blood samples. This is due to limitations in the available technology as well as sample limitations.
-
- Terminated Z-2732-2016
There is a possibility of specimen misidentification for the stand-alone DxH 800 and DxH 600 where the Primary Identifier is configured to Tube Position ID.
-
- Terminated Z-2692-2016
Beckman Coulter is recalling the UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System to reduce the potential for flammable liquids coming into contact with electrical components in the Stainer module possibly causing a fire.
-
- Terminated Z-0432-2016
- Terminated Z-0434-2016
- Terminated Z-0433-2016
Beckman Coulter is recalling the DxH 800 Slidemaker Stainer Software version 3.0.2.0 and DxH 600 software version 1.1.1.0 because the software for the DxH systems allows the creation of multiple orders with the same Specimen identification (ID) but different Patient identification when manually editing pending orders at the System Manager.
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
- European Union
- United States
Specification
- Trade Name
- UniCel® DxH 900 Coulter® Cellular Analysis System EUDAMEDDxH 900 Hematology System GUDID
- Generic Name
- Haematological cell analyser IVD
- Model / Ref / Catalog
- C11478
- Primary DI / UDI-DI
- D-C11478UC EUDAMED15099590701574 GUDID
- Basic UDI-DI
- B-C11478UC
- FDA Product Code
- GKZ
- EMDN Code
- W0202010101 AUTOMATED WHOLE BLOOD HEMATOLOGY ANALYSERS - LOW ROUTINE (Throughput <= 100 samples/hr)
- GMDN Term
- Haematological cell analyser IVD
- Device Class
- Class IVD General EUDAMEDClass II GUDID
- Regulation Number
- 864.5220
- EU market of registration
- Ireland
- GUDID distribution status
- In Commercial Distribution
Truth per registry
| Field | EUDAMED | GUDID |
|---|---|---|
| Primary DI / UDI-DI | D-C11478UC | 15099590701574 |
| Device Class | Class IVD General | Class II |
| Sterile | No | Yes |
Other devices from Beckman Coulter, Inc.
Properties
Other properties (17)
- Single Use: No
- Implantable: No
- Contains Latex: No
- Human Tissues: No
- Animal Tissues: No
- Microbial Substances: No
- Reagent: No
- Self Testing: No
- Near-Patient Testing: No
- Companion Diagnostics: No
- Over-the-Counter: No
- Convenience Kit: No
- Combination Product: No
- HCT/P: No
- Premarket Exempt: No
- Lot / Batch Number: No
- Expiration Date: No