Skip to content

Dynagen™ EL ICD VR D151

by Boston Scientific

Dynagen EL ICD VR D151 is a single-chamber implantable defibrillator (FDA Class III; FDA product code LWS, Implantable cardioverter defibrillator (non-CRT)) made by Boston Scientific, approved under FDA PMA P960040 on 18 Jul 1997 and listed in FDA GUDID since 6 Nov 2014 (UDI-DI 00802526534300), active as of 22 Oct 2026. 3 FDA recalls are linked to it.

Active United States. Also listed in GUDID (US).

PMA

Valid until

No expiry

PMA approval is not time-limited

FDA recalls

1 ongoing

This device is named in the recall

Documents

PMA P960040

FDA submission, from GUDID

Trust

Unverified

Source GUDID, synced 22 Oct 2026.

Regulatory status

as of registry sync 22 Oct 2026

RegionAuthorizationNumberDateStatus
United StatesGUDID listing Number00802526534300 UDI-DI Date6 Nov 2014ListedStatusListed GUDID: in commercial distribution
United StatesFDA PMAAPPR from GUDID Approved NumberP960040 Date18 Jul 1997DecidedStatusActive

Official records

From the FDA record

Device description
Implantable Cardioverter Defibrillator VR
MRI safety
MR Conditional
Natural rubber latex
No latex statement on the label
Sterility
Supplied sterile

From the GUDID record (AccessGUDID), as read on 22 Oct 2026, as the labeller entered it.

Across registries

2 registries list this device

  • FDA PMA P960040 decided 18 Jul 1997 Active
  • FDA GUDID UDI-DI 00802526534300 listed 6 Nov 2014 Listed

Recalls

14 recalls · 1 ongoing · FDA

    • Ongoing Z-1670-2022 FDA Class II initiated 7 Jul 2022

    There is potential for unanticipated software anomalies to occur when using the Model 2868 Application loaded on the 3120 ZOOM LATITUTDE programmer when communicating with certain Boston Scientific transvenous defibrillators.

  • 3 recalls, same reason

    • Terminated Z-0077-2018 FDA Class II initiated 4 Oct 2017
    • Terminated Z-0078-2018 FDA Class II initiated 4 Oct 2017
    • Terminated Z-0079-2018 FDA Class II initiated 4 Oct 2017 on shared PMA P960040

    The devices have an incorrect firmware configuration.

    • Terminated Z-1555-2010 FDA Unclassified (pre-2012) initiated 15 Mar 2010 on shared PMA P960040

    Boston Scientific Corporation has stopped shipment and is retrieving unimplanted devices from US hospital shelves of all its implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds) after determining that some manufacturing process changes were not submitted for approval to the US Food & Drug Administration.

    • Terminated Z-0610-2010 FDA Unclassified (pre-2012) initiated 1 Dec 2009 on shared PMA P960040

    BSC CRM provided important patient management information regarding Boston Scientific COGNIS CRT-Ds and TELIGEN ICDs. Engineering simulations and field reports indicate that in some cases, implanted devices may encounter sufficient mechanical stress to weaken the bond between the header and case when positioned subpectorally. A weakened header bond may alter lead impedance and introduce noise

    • Terminated Z-1621-2009 FDA Unclassified (pre-2012) initiated 4 Mar 2009 on shared PMA P960040

    In April 2007, Boston Scientific CRM communicated with physicians regarding the potential for reduced Elective Replacement Indicator (ERI) to Battery End of Life (EOL) time in a subset of implantable defibrillators due to degradation of a low-voltage capacitor. Since that time, the April 2007 advisory population has not experienced any clinically significant changes to either the rate of occurrenc

  • 2 recalls, same reason

    • Terminated Z-1513-2008 FDA Unclassified (pre-2012) initiated 7 Jan 2008 on shared PMA P960040
    • Terminated Z-1514-2008 FDA Unclassified (pre-2012) initiated 7 Jan 2008 on shared PMA P960040

    Repetitive mechanical stress applied to the titanium case can induce component damage and device malfunction in certain ICD and CRT-D devices implanted subpectorally with the serial number facing the ribs. Devices implanted subcutaneously or in a subpectoral position with the serial number facing away from the ribs are not included in this advisory. This update identifies additional VITALITY ICD

  • 2 recalls, same reason

    • Terminated Z-0050-2008 FDA Unclassified (pre-2012) initiated 16 Mar 2007 on shared PMA P960040
    • Terminated Z-0051-2008 FDA Unclassified (pre-2012) initiated 16 Mar 2007 on shared PMA P960040

    Atypical Charge Time Behavior: 1)-End of Life (EOL) or ERI is displayed during mid-life (typically 24-48 months), even though battery capacity remains available. 2)- Extended ERI charge time limits: Charge times during mid-life may remain below a normal extension of the ERI charge time . ****** Prophylactic replacement of these devices prior to ERI is not recommended. As stated in the Product U

    • Terminated Z-1297-06 FDA Unclassified (pre-2012) initiated 24 Jun 2006 on shared PMA P960040

    Guidant has determined that low-voltage capacitors from a single component supplier may malfunction in a manner that can lead to premature battery depletion in the affected devices.

    • Terminated Z-1292-06 FDA Unclassified (pre-2012) initiated 15 May 2006 on shared PMA P960040

    Guidant has identified a pattern of premature battery depletion in devices manufactured with a single lot of capacitors from a single supplier. If function of this low voltage capacitor is compromised, the battery may deplete prematurely.

    • Terminated Z-0828-06 FDA Unclassified (pre-2012) initiated 7 Apr 2006 on shared PMA P960040

    77 distributed devices may have a cracked layer of insulation in a flexible hybrid circuit, leading to internal shorting during charging such as during routine capacitor formation or for defibrillation. This would likely render the device non-functional, resulting in no output and/or telemetry communications.

Recalls on a shared submission name another device cleared under the same FDA submission as this one.

Timeline

16 dated events

  1. Recall initiated FDA recalls Z-1513-2008
  2. Recall initiated FDA recalls Z-1514-2008
  3. Recall initiated FDA recalls Z-1621-2009
  4. Recall initiated FDA recalls Z-0610-2010
  5. Recall initiated FDA recalls Z-1555-2010
  6. Listed FDA GUDID 00802526534300
  7. Recall initiated FDA recalls Z-0077-2018
  8. Recall initiated FDA recalls Z-0078-2018
  9. Recall initiated FDA recalls Z-0079-2018
  10. Recall initiated FDA recalls Z-1670-2022

6 earlier events are in the sections above.

Coverage

1 jurisdiction

United States of America
  • Active
  • Expiring
  • Withdrawn
  • Listed
  • Not covered

United States — FDA PMA Active

Specification

GUDID

Trade name
Dynagen™ El Icd Vr
Generic name
Single-chamber implantable defibrillator
Model / Ref / Catalog
D151
Primary DI / UDI-DI
00802526534300
PMA number
FDA product code
LWS
GMDN term
Single-chamber implantable defibrillator
Device class
Class III
GUDID distribution status
In Commercial Distribution
MRI safety
MR Conditional

Similar devices

Other devices filed under FDA product code LWS (Implantable cardioverter defibrillator (non-CRT)).

Other devices from Boston Scientific

See all devices from Boston Scientific →

Properties

  • Supplied sterile
  • Single use
  • Implantable
  • Prescription use
  • Serial number
  • Expiration date
Other properties (9, all answered No)