Skip to content

E100 Series SPO2 Sensors

by Epic Medical Equipment Services, Inc. 510(k)

Active United States

510(k) K970098

Regulatory status

as of 29 Oct 2026

RegionAuthorizationNumberDateStatus
United StatesFDA 510(k)SESE Substantially Equivalent · Traditional NumberK970098 Date30 Apr 1998ClearedStatusActive

Coverage

1 jurisdiction

United States of America
  • Active
  • Expiring
  • Withdrawn
  • Listed
  • Not covered

United States — FDA 510(k) Active

Recalls

1 recall · FDA

    • Terminated Z-1188-2008 FDA Unclassified (pre-2012) initiated 18 May 2007

    Incompatible components: The Finger Clip Sensors - with a Serial Number prefix of PA514, PA515, YE508, ZC502 and ZG512 - are missing an additional diode and higher resistor in its existing configuration making them inoperable when used to interface with those Ohmeda Tuffsat monitors previously re-configured by Ohmeda to include a detection circuit for diodes.

Specification

510(k)

510(k) Number
FDA Product Code
DQA
Device Class
Class II
Regulation
21 CFR Part 870.2700
Advisory Committee
Cardiovascular
FDA Decision
SESE Substantially Equivalent

Other devices from Epic Medical Equipment Services, Inc.

See all devices from Epic Medical Equipment Services, Inc. →