Ecat Exact & Ecat Exact Hr+
by Cti Pet Systems, Inc. 510(k)
Active
510(k) K962797
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK962797 | Date15 Oct 1996 | StatusActive |
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not known
- Not covered
United States — FDA 510(k) Active
Recalls
-
- Terminated Z-0248-2008
- Terminated Z-0249-2008
- Terminated Z-0250-2008
- Terminated Z-0247-2008
- Terminated Z-0246-2008
- Terminated Z-0251-2008
- Terminated Z-0252-2008
Visual asymmetric appearance; An asymmetry can be introduced into the attenuation-corrected PET images when using iterative reconstruction in combination with a specific number of subsets and reconstruction matrix sizes. (This problem only occurs when using protocols other than the default Biograph protocols)
-
- Terminated Z-0269-2008
- Terminated Z-0270-2008
- Terminated Z-0272-2008
- Terminated Z-0271-2008
Incorrect Decay Correction Results - ECAT System's software versions 7.2.2 and 7.4 calculate the decay correction for emission data incorrectly, while performing whole-body studies with all transmissions followed by all emissions.
-
- Terminated Z-0273-2008
Incorrect Data Results - ECAT System's software versions 7.2.2 and 7.4 may calculate the decay correction for emission data incorrectly while performing whole-body studies with all transmissions followed by all emissions.
-
- Terminated Z-0819-05
Failure to properly secure the Mobile Bed Supports during transit could cause the support screws on the Patient Handling System to become loose and possibly break due to increased vibration during transit.
-
- Terminated Z-0674-03
Incorrect patient information is displayed in the reconstructed images of a whole body patient scan.
Specification
- 510(k) Number
- FDA Product Code
- KPS
- Device Class
- Class II
- Regulation
- 21 CFR Part 892.1200
- Advisory Committee
- Radiology
- FDA Decision
- SESE Substantially Equivalent