Eg-3670URK
by Pentax Medical Company 2 variants
510(k) K061068
Active
510(k)
Valid until
No expiry
FDA recalls
None ongoingDocuments
Trust
Unverified
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK061068 | Date3 May 2006 | StatusActive |
Recalls
-
- Terminated Z-3067-2017
Error in reprocessing instructions in Revisions 13 and 14 of IFU Z845, used with EG-3670URK and EG-3870UTK ultrasound gastroscopes. In the disinfection step, the IFU has a warning indicating the cleaning detergent solution should remain in contact with the all internal channels and external endoscope surfaces. The disinfecting solution, not the cleaning detergent, should be used for this step.
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- 510(k) Number
- FDA Product Code
- ODG
- Device Class
- Class II
- Regulation
- 21 CFR Part 876.1500
- Advisory Committee
- Gastroenterology, Urology
- FDA Decision
- SESE Substantially Equivalent