Eg-3870UTK

by Pentax Medical Company

510(k) K090197

Active United States

510(k)

Valid until

No expiry

510(k) clearance is not time-limited

FDA recalls

None ongoing

2 closed

Documents

Trust

Unverified

Registry data, not yet checked by deviCERTS. Source FDA 510(k), synced 10 Oct 2026. What this means

Regulatory status

as of registry sync 10 Oct 2026

RegionAuthorizationNumberDateStatus
United StatesFDA 510(k)SESE Substantially Equivalent · Traditional NumberK090197 Date19 May 2009ClearedStatusActive

Recalls

2 recalls · FDA

    • Terminated Z-3068-2017 FDA Class II initiated 23 May 2017

    Error in reprocessing instructions in Revisions 13 and 14 of IFU Z845, used with EG-3670URK and EG-3870UTK ultrasound gastroscopes. In the disinfection step, the IFU has a warning indicating the cleaning detergent solution should remain in contact with the all internal channels and external endoscope surfaces. The disinfecting solution, not the cleaning detergent, should be used for this step.

    • Terminated Z-1489-2012 FDA Not yet classified initiated 14 Mar 2012 on shared 510(k) K090197

    Endoscope ultrasound balloon evacuation tube may become clogged with patient body fluid if users fail to properly follow instructions for pre-procedure check, operating the suction valve during procedure and pre-cleaning procedure.

Recalls on a shared submission name another device cleared under the same FDA submission as this one.

Coverage

1 jurisdiction

United States of America
  • Active
  • Expiring
  • Withdrawn
  • Listed
  • Not covered

United States — FDA 510(k) Active

Specification

510(k)

510(k) Number
FDA Product Code
ODG
Device Class
Class II
Regulation
21 CFR Part 876.1500
Advisory Committee
Gastroenterology, Urology
FDA Decision
SESE Substantially Equivalent