Skip to content

Eia Smdp Immunoassay

by Phadia US, Inc.

510(k) K132631

Active United States

510(k)

Valid until

No expiry

510(k) clearance is not time-limited

FDA recalls

None ongoing

1 closed

Documents

Trust

Unverified

Registry data, not yet checked by deviCERTS. Source FDA 510(k), synced 11 Oct 2026. What this means

Regulatory status

as of registry sync 11 Oct 2026

RegionAuthorizationNumberDateStatus
United StatesFDA 510(k)SESE Substantially Equivalent · Traditional NumberK132631 Date29 Aug 2014ClearedStatusActive

Recalls

1 recall · FDA

    • Terminated Z-0697-2016 FDA Class II initiated 28 Dec 2015

    All equivocal and positive results (greater or equal 7 EliA U/ml) on EliA SmDP well lot 0018 may be incorrect and must be considered invalid. Patient samples in complaint investigations caused unspecific signals up to 22 U/ml on EliA SmDP well lot 0018. The nonspecific signals are not caused by anti-Sm antibodies, nor by streptavidin antibodies.

Coverage

1 jurisdiction

United States of America
  • Active
  • Expiring
  • Withdrawn
  • Listed
  • Not covered

United States — FDA 510(k) Active

Specification

510(k)

510(k) Number
FDA Product Code
LKP
Device Class
Class II
Regulation
21 CFR Part 866.5100
Advisory Committee
Immunology
FDA Decision
SESE Substantially Equivalent

Other devices from Phadia Us Inc.