Ekosonic Mach 4 Endovascular Device/Ekosonic Control Unit
510(k) K140151
Active
Valid until
No expiry
FDA recalls
None ongoingDocuments
Trust
Unverified
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK140151 | Date21 May 2014 | StatusActive |
Recalls
-
- Terminated Z-2854-2016
The recalling firm identified ten devices which were distributed and could potentially be non-compliant. These devices potentially underwent a routine manufacturing test with non-compliant parameters, which may result in a small blister in the epoxy over transducers in the treatment zone. It is not known at the time the recall was initiated whether the non-compliant test parameters, or presence of the blisters, adversely affect operation of the device.
-
- Terminated Z-0864-2016
The EkoSonic kit, including the individually labeled MicroSonic Device (MSD) and Intelligent Drug Delivery Catheter (IDDC) are labeled with a 135 cm working length, and programmed with a 135 cm family code, however the MSD actual working length is 106 cm.
-
- Terminated Z-2148-2015
The Connector Interface Cable (CIC) was not recognized by the EkoSonic PT-3B control unit.
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- 510(k) Number
- FDA Product Code
- QEY
- Device Class
- Class II
- Regulation
- 21 CFR Part 870.5150
- Advisory Committee
- Cardiovascular
- FDA Decision
- SESE Substantially Equivalent