Ektachem Clinical Chemistry Slide Glucose (Glu)
by Johnson & Johnson Clinical Diagnostics, Inc.
510(k) K955286
Active
Valid until
No expiry
FDA recalls
2 ongoingDocuments
Trust
Unverified
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK955286 | Date17 May 1996 | StatusActive |
Recalls
-
- Ongoing Z-2560-2025
- Ongoing Z-2559-2025
VITROS Chemistry Products LAC Slides on VITROS XT 3400 and VITROS XT 7600 Systems are experiencing an increased occurrence of calibration failure related to lots from coating 0130 and above with Condition Code TH4-63J during calibration, causing the delay of lac test results and potentially impact patient management. The issue was identified by QuidelOrtho through an increase in customer complaints compared to the historical baseline with average failure rate of 1% for coating 0130 in 2025 while average of failure rate for other coating was 0.06% in 2025 and average of failure rate for all coatings is 0.03% for 2024.
-
- Terminated Z-2202-2019
Potentially Biased Results using VITROS Chemistry Products GLU Slides
-
- Terminated Z-0090-2017
Ortho Clinical Diagnostics confirmed that N-Acetylcysteine (NAC), a drug used for acetaminophen overdose, mucolytic therapy and is used as nutritional supplement, can interfere with VITROS GLU, LAC, TRIG, and URIC Slides assays.
-
- Terminated Z-0089-2017
Ortho Clinical Diagnostics confirmed that N-Acetylcysteine (NAC), a drug used for acetaminophen overdose, mucolytic therapy and is used as nutritional supplement, can interfere with VITROS GLU, LAC, TRIG, and URIC Slides assays. Ortho also tested several compounds following industry standard CLSI EP7-A2 for potential interference on VITROS GLU Slides during validation testing for a proposed change and confirmed Dextran 40 and Glutathione may interfere with current VITROS GLU Slides.
-
- Terminated Z-0314-2014
Ortho Clinical Diagnostics is recalling Slide Dispense cartridges used with some VITROS Chemistry Product Slides due to slide dispense issues.
-
- Terminated Z-1224-03
Results of glucose measurements on sodium fluoride/potassium oxalate specimens can be negatively biased when using Vitros GLU and GLU DT slides stored in the refrigerator. GLU slides are processed by high volume chemistry systems typically used in hospital laboratories.
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- 510(k) Number
- FDA Product Code
- CGA
- Device Class
- Class II
- Regulation
- 21 CFR Part 862.1345
- Advisory Committee
- Clinical Chemistry
- FDA Decision
- SESE Substantially Equivalent