Elecsys 2010 Analyzer

by Boehringer Mannheim Corp.

510(k) K961481

Active United States

510(k)

Valid until

No expiry

510(k) clearance is not time-limited

FDA recalls

None ongoing

6 closed

Documents

Trust

Unverified

Registry data, not yet checked by deviCERTS. Source FDA 510(k), synced 22 Oct 2026. What this means

Regulatory status

as of registry sync 22 Oct 2026

RegionAuthorizationNumberDateStatus
United StatesFDA 510(k)SESE Substantially Equivalent · Traditional NumberK961481 Date14 Jun 1996ClearedStatusActive

Recalls

6 recalls · FDA

    • Terminated Z-0499-2018 FDA Class II initiated 5 Jul 2017

    A software malfunction can occur on the cobas e 411 and Elecsys 2010 analyzers in the Sample& Control data file which may lead to a potential data mismatch.

    • Terminated Z-0095-2017 FDA Class II initiated 15 Jul 2016

    Roche has received four reports of injury caused by the sharp edge of the center guide rail of an analyzers sample rack tray. The cleaning procedure for the sample rack tray has been revised to clarify the cleaning process.

    • Terminated Z-1790-2014 FDA Class II initiated 21 May 2014

    Under certain conditions loose ProCell/CleanCell aspiration tube filters may cause: consecutive discrepant results; sipper alarm; foam on reservoirs and system reagent alarms.

    • Terminated Z-1218-2012 FDA Not yet classified initiated 31 Jan 2012

    Elecsys 2010 customers have reported getting no value and flag 48 results below expected value range after performing automatic dilution for a sample that previously recovered above the measuring range. Incorrect but believable results may occur for the CEA, Rubella IgG or SHBG assays. The other affected assays will generate either no value or a negative result, both of which are not bel

    • Terminated Z-0165-2009 FDA Unclassified (pre-2012) initiated 24 Jun 2008

    A software bug may result in pipetting from an incorrect reagent pack and/or assigning calibration curve parameters incorrectly.

    • Terminated Z-0868-2008 FDA Unclassified (pre-2012) initiated 21 Mar 2008

    Possible mismatch between patient and result.

Coverage

1 jurisdiction

United States of America
  • Active
  • Expiring
  • Withdrawn
  • Listed
  • Not covered

United States — FDA 510(k) Active

Specification

510(k)

510(k) Number
FDA Product Code
JJE
Device Class
Class I
Regulation
21 CFR Part 862.2160
Advisory Committee
Clinical Chemistry
FDA Decision
SESE Substantially Equivalent