Elekta Oncology Systems Precise Treatment Table
510(k) K983678
Active
510(k)
Valid until
No expiry
FDA recalls
None ongoingDocuments
Trust
Unverified
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK983678 | Date15 Jan 1999 | StatusActive |
Recalls
-
- Terminated Z-1238-2015
It is possible to position the Treatment Table with errors greater than 5mm and no inhibits are displayed, if the positional sensors are not correctly installed during corrective maintenance. Clinical mistreatment may occur if a position error is not detected.
-
- Terminated Z-1072-2013
The Column rotation can accidentally be moved to a different position which in the locked position.
-
- Terminated Z-1030-2011
Possible error in the table longitudinal position.
-
- Terminated Z-1432-2010
In some instances, the bolts which are supplied with the table have sheared off, resulting in inability to operate the powered table.
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- 510(k) Number
- FDA Product Code
- JAI
- Device Class
- Class II
- Regulation
- 21 CFR Part 892.5770
- Advisory Committee
- Radiology
- FDA Decision
- SESE Substantially Equivalent