Elekta Unity
510(k) K182076
Elekta Unity is a medical device (FDA Class II; FDA product code IYE, Accelerator, Linear, Medical) made by Elekta Limited, cleared under FDA 510(k) K182076 on 4 Dec 2018, active as of 22 Oct 2026. 4 FDA recalls are linked to it.
Active
Valid until
No expiry
FDA recalls
None ongoingDocuments
Trust
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK182076 | Date4 Dec 2018 | StatusActive |
Official records
- FDA 510(k) database K182076
From the FDA record
- Received by the FDA
- 1 Aug 2018
- FDA decision
- 4 Dec 2018 (Substantially Equivalent)
- FDA review time
- 125 days from receipt to decision
- Public document
- 510(k) summary on file with the FDA
Recalls
-
- Terminated Z-1980-2019
The QA software solution to perform the MR to MV alignment check, does not display the stored MR to MV offset values. The user is unable to independently inspect the values during their QA.
-
- Terminated Z-1297-2019
There is no warning in the Elekta Unity manual for the administration of gadolinium-based or other contrast agents for imaging or image guidance purposes has not been validated.
-
- Terminated Z-1109-2019
Users need to be aware when using these protocols for daily on line plan adaptation that: 1) The images acquired using these protocols do not represent the average position of the anatomy during the respiratory motion cycle. The images are based on data acquired around full expiration. 2) The display of the images in the Elekta Unity Application software does not provide information about the protocol used to acquire the image eg. with or without respiratory triggering. Users must select an appropriate scan protocol that is representative of the respiratory phase used in the reference plan.
-
- Completed Z-0706-2019
In rare circumstances, due to reasons such as inadequate hospital facility maintenance, water leakage, and weakening of speaker fixings during the lifetime of the system, the speakers can become dislodged and be pulled into the bore.
Timeline
- Cleared FDA 510(k) K182076
- Recall initiated FDA recalls Z-0706-2019
- Recall initiated FDA recalls Z-1109-2019
- Recall initiated FDA recalls Z-1297-2019
- Recall initiated FDA recalls Z-1980-2019
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- 510(k) Number
- FDA product code
- IYE
- Device class
- Class II
- Regulation
- 21 CFR Part 892.5050
- Advisory committee
- Radiology
- FDA decision
- SESE Substantially Equivalent
Devices cleared under K182076
Devices that FDA's GUDID lists as cleared or approved under this submission number. One submission usually covers a whole product family, so the sizes and models are listed here once.
| UDI-DI | Size and differences | |||
|---|---|---|---|---|
| Elekta Unity 1059073 | 1059073 | 05060191071420 | — | On the market |
| Elekta Unity 1065120 | 1065120 | 05060191071437 | — | On the market |
| Elekta Unity 1536549 | 1536549 | 05060191071321 | — | On the market |
Similar devices
Other devices filed under FDA product code IYE (Accelerator, Linear, Medical).
- COREGISTRATION FUNCTION by Nomos Corp.
- LINAC-BASED STEROTACTIC RADIOSURGERY by Radionics Software Applications, Inc.
- Cranial 4Pi Immobilization by Brainlab AG
- MODIFICATION TO: ON-BOARD IMAGER DEVICE by Varian Medical Systems, Inc.
- KLARITY MEDICAL R550-CAEP by Klarity Medical Products, LLC
- GILL-THOMAS-COSMAN(GTC) DENTAL TRAY by Radionics Software Applications, Inc.
- MOSAIQ Oncology Information System by ELEKTA SOLUTIONS AB
- Raycast High Precision Lung Board Solution by Orfit Industries NV
- Klarity Medical R550-HS1 by Klarity Medical Products, LLC
- VARIAN CLINAC 1800S by Varian Assoc., Inc.