  "format": "devicerts.device.v1",
  "cite_as": "Elekta Unity, Elekta Limited. deviCERTS medical device catalogue. https://devicerts.com/devices/elekta-unity-iye-k182076 (data as of 2026-11-20).",
  "source": {
    "canonical_url": "https://devicerts.com/devices/elekta-unity-iye-k182076",
    "json_url": "https://devicerts.com/devices/elekta-unity-iye-k182076.json",
    "markdown_url": "https://devicerts.com/devices/elekta-unity-iye-k182076.md",
    "publisher": "deviCERTS (devicerts.com)",
    "data_sources": [
      {
        "registry": "FDA 510(k)",
        "url": "https://open.fda.gov/apis/device/510k/"
      },
      {
        "registry": "FDA recalls",
        "url": "https://open.fda.gov/apis/device/recall/"
      }
    ],
    "registry_records_last_synced": "2026-10-22",
    "status_as_of": "2026-11-20",
    "license": "CC BY 4.0",
    "license_url": "https://creativecommons.org/licenses/by/4.0/",
    "license_scope": "The deviCERTS layout and text are licensed under CC BY 4.0. Registry data remain under their source terms: U.S. FDA data (openFDA, AccessGUDID) are U.S. public domain; EUDAMED data are subject to the European Commission reuse terms.",
    "attribution": "DevicErts, https://devicerts.com/devices/elekta-unity-iye-k182076",
    "disclaimer": "Compiled by deviCERTS (devicerts.com) from public medical device registries (EUDAMED, European Commission; U.S. FDA via openFDA; AccessGUDID, U.S. National Library of Medicine). deviCERTS is not an official register and gives no regulatory advice: the record in the registry itself prevails."
  },
  "device": {
    "name": "Elekta Unity",
    "registry_name": "Elekta Unity",
    "url": "https://devicerts.com/devices/elekta-unity-iye-k182076",
    "manufacturer": {
      "name": "Elekta Limited",
      "url": "https://devicerts.com/manufacturers/elekta-oy",
      "country": "Finland",
      "country_basis": null,
      "fda_registration_country": null
    },
    "status": {
      "state": "active",
      "label": "Active",
      "detail": "United States"
    },
    "classification": [
      {
        "class": "Class II",
        "system": "US FDA"
      }
    ],
    "identifiers": [
      {
        "label": "510(k) Number",
        "values": [
          {
            "value": "K182076",
            "registries": [
              "FDA 510(k)"
            ]
          }
        ]
      }
    ],
    "product_codes": [
      "IYE"
    ],
    "emdn": [],
    "gmdn": [],
    "description": null,
    "noindex": false
  },
  "authorizations": [
    {
      "registry": "510(k)",
      "region": "United States",
      "type": "FDA 510(k)",
      "number": "K182076",
      "number_label": null,
      "device_udi_di": null,
      "date": "2018-12-04",
      "date_kind": "decision",
      "decision_code": "SESE",
      "expires": null,
      "withdrawn": null,
      "status": "Active",
      "status_note": null,
      "derived_from_gudid": false
    }
  ],
  "no_longer_listed_by": [],
  "certificates": [],
  "recalls": [
    {
      "number": "Z-1980-2019",
      "classification": "Class II",
      "status": "terminated",
      "initiated": "2019-06-24",
      "reason": "The QA software solution to perform the MR to MV alignment check, does not display the stored MR to MV offset values. The user is unable to independently inspect the values during their QA.",
      "applies": "this device",
      "via": "K182076"
    },
    {
      "number": "Z-1297-2019",
      "classification": "Class II",
      "status": "terminated",
      "initiated": "2019-04-03",
      "reason": "There is no warning in the Elekta Unity manual for the administration of gadolinium-based or other contrast agents for imaging or image guidance purposes has not been validated.",
      "applies": "this device",
      "via": "K182076"
    },
    {
      "number": "Z-1109-2019",
      "classification": "Class II",
      "status": "terminated",
      "initiated": "2019-03-12",
      "reason": "Users need to be aware when using these protocols for daily on line plan adaptation that:\n1) The images acquired using these protocols do not represent the average position of the anatomy during the respiratory motion cycle. The images are based on data acquired around full expiration.\n2) The display of the images in the Elekta Unity Application software does not provide information about the protocol used to acquire the image eg. with or without respiratory triggering.\nUsers must select an appropriate scan protocol that is representative of the respiratory phase used in the reference plan.",
      "applies": "this device",
      "via": "K182076"
    },
    {
      "number": "Z-0706-2019",
      "classification": "Class II",
      "status": "completed",
      "initiated": "2019-01-04",
      "reason": "In rare circumstances, due to reasons such as inadequate hospital facility maintenance, water leakage, and weakening of speaker fixings during the lifetime of the system, the speakers can become dislodged and be pulled into the bore.",
      "applies": "this device",
      "via": "K182076"
    }
  ],
  "eu_market_availability": null,
  "properties": [
    {
      "group": "Identifiers",
      "fields": [
        {
          "label": "510(k) Number",
          "values": [
            {
              "value": "K182076",
              "registries": [
                "FDA 510(k)"
              ]
            }
          ]
        },
        {
          "label": "FDA product code",
          "values": [
            {
              "value": "IYE",
              "registries": [
                "FDA 510(k)"
              ]
            }
          ]
        }
      ]
    },
    {
      "group": "Classification",
      "fields": [
        {
          "label": "Device class",
          "values": [
            {
              "value": "Class II",
              "registries": [
                "FDA 510(k)"
              ]
            }
          ]
        },
        {
          "label": "Regulation",
          "values": [
            {
              "value": "21 CFR Part 892.5050",
              "registries": [
                "FDA 510(k)"
              ]
            }
          ]
        },
        {
          "label": "Advisory committee",
          "values": [
            {
              "value": "Radiology",
              "registries": [
                "FDA 510(k)"
              ]
            }
          ]
        },
        {
          "label": "FDA decision",
          "values": [
            {
              "value": "SESE",
              "registries": [
                "FDA 510(k)"
              ]
            }
          ]
        }
      ]
    }
  ]
}