Elia Ccp, Model 14-5515-01; Elia Ccp Control, Model 83-1009-01
by Sweden Diagnostics (Us), Inc.
510(k) K061165
Active
510(k)
Valid until
No expiry
FDA recalls
None ongoingDocuments
Trust
Unverified
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK061165 | Date3 Aug 2006 | StatusActive |
Recalls
-
- Terminated Z-1253-2017
Erroneous EliA test results related to problems with liquid level detection and diluent pipetting in Phadia 250 instruments, and deformed bottles. Investigation showed problems with the liquid level detection in diluent bottle batch JJCA resulting in aspiration of no or too low volumes of liquid from the diluent bottle and from the dilution plate.
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- 510(k) Number
- FDA Product Code
- NHX
- Models / Variants
- Device Class
- Class II
- Regulation
- 21 CFR Part 866.5775
- Advisory Committee
- Immunology
- FDA Decision
- SESE Substantially Equivalent