Empira Rx Ptca Dilatation Catheter, Empria Nc Rx Ptca Dilatation Catheter
510(k) K110133
Active
510(k)
Valid until
No expiry
FDA recalls
None ongoingDocuments
Trust
Unverified
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK110133 | Date10 Jun 2011 | StatusActive |
Recalls
-
- Terminated Z-2556-2014
- Terminated Z-2557-2014
The balloon may exhibit resistance to the removal of the balloon cover. Weakening of the affected product in the region of the proximal balloon seal may occur when excessive force is applied to remove the balloon cover. This may lead to balloon burst, shaft separation or deflation difficulties, which could result in procedural delay, vessel trauma or surgery.
-
- Terminated Z-1784-2014
Dilatation catheters could exhibit radial versus axial tears should they burst during inflation.
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- 510(k) Number
- FDA Product Code
- LOX
- Device Class
- Class II
- Regulation
- 21 CFR Part 870.5100
- Advisory Committee
- Cardiovascular
- FDA Decision
- SESE Substantially Equivalent