Skip to content

EndoWrist 470530

by Intuitive Surgical, Inc.

Ref 470530

EndoWrist 470530 is a robotic surgical cutting stapler (EU MDR Class IIa; FDA Class II; FDA product code NAY, System,surgical,computer controlled instrument) made by Intuitive Surgical, Inc., registered in EUDAMED on 4 Jun 2026 (Basic UDI-DI 08868741209658X) and listed in FDA GUDID since 14 Jul 2016 (UDI-DI 00886874112717), active as of 22 Oct 2026. 2 FDA recalls are linked to it.

Active European Union. Also listed in GUDID (US).

CE MDR

Valid until

No date given

EUDAMED holds no validity date

FDA recalls

None ongoing

2 closed

Documents

None linked in EUDAMED

A group certificate may exist

Trust

Unverified

Source EUDAMED, GUDID, synced 22 Oct 2026.

Regulatory status

as of registry sync 22 Oct 2026

RegionAuthorizationNumberDateStatus
European UnionCE/MDR EUDAMED Number08868741209658X Basic UDI-DI 00886874112717UDI-DIDate4 Jun 2026RegisteredStatusActive
United StatesGUDID listing GUDID Number00886874112717 UDI-DI Date14 Jul 2016ListedStatusListed GUDID: in commercial distribution

Official records

From the FDA record

Device description
CURVED-TIP STAPLER 30
MRI safety
Labeling does not contain MRI Safety Information
Natural rubber latex
No latex statement on the label
Sterility
Supplied non-sterile; sterilise before use (Moist Heat or Steam Sterilization)

From the GUDID record (AccessGUDID), as read on 22 Oct 2026, as the labeller entered it.

Across registries

2 registries list this device

  • EUDAMED Basic UDI-DI 08868741209658X registered 4 Jun 2026 Active
  • FDA GUDID UDI-DI 00886874112717 listed 14 Jul 2016 Listed

Recalls

2 recalls · FDA

    • Terminated Z-2350-2016 FDA Class II initiated 14 Jul 2016 catalog no. 470530 listed

    An increase in the number of da Vinci EndoWrist Stapler complaints where the customer experienced a Stapler Firing Failed message as well as review of da Vinci Xi Surgical System logs found an increased trend

    • Terminated Z-2677-2016 FDA Class II initiated 14 Jul 2016 catalog no. 470530 listed

    Field failures were reported related to the da Vinci Xi Surgical System EndoWrist Stapler 45 and 30 instruments having an interruption in shaft rotation and / or instrument jaws not opening.

A catalog-number match means the recall lists this device's catalog or model number; the software version and lot are not checked.

Timeline

4 dated events

  1. Listed FDA GUDID 00886874112717
  2. Recall initiated FDA recalls Z-2350-2016
  3. Recall initiated FDA recalls Z-2677-2016
  4. Registered EUDAMED 08868741209658X

Coverage

2 jurisdictions

United States of AmericaSwedenFinlandLatviaEstoniaDenmarkLithuaniaIrelandNetherlandsGermanyPolandFranceBelgiumLuxembourgCzech RepublicSlovakiaAustriaHungaryRomaniaSloveniaBulgariaPortugalSpainItalyCroatiaGreeceMaltaCyprus
  • Active
  • Expiring
  • Withdrawn
  • Listed
  • Not covered
  • European Union CE/MDR Active
  • United States GUDID listing Listed

Specification

EUDAMED · GUDID

Trade name
EndoWrist Stapler 30 Curved-Tip EUDAMED
EndoWrist GUDID
Generic name
Robotic surgical cutting stapler, reusable
Device name
EndoWrist Stapler instrument
Model / reference
470530
Primary DI / UDI-DI
00886874112717
Basic UDI-DI
08868741209658X
FDA product code
NAY
EMDN code
H02030199 Linear staplers for videosurgery - other
GMDN term
Robotic surgical cutting stapler, reusable
Device class
Class IIa EUDAMED
Class II GUDID
Regulation number
876.1500
EU registration type
Standard device
EU market of registration
Germany
GUDID distribution status
In Commercial Distribution
Sterilisation method
Moist Heat or Steam Sterilization
Size
Outer Diameter — 12 Millimeter

Truth per registry

2 sources · 1 field differs

FieldEUDAMEDGUDID
Device classEUDAMED
Class IIa
GUDID
Class II

Device class is stated on each registry's own scale: Class IIa in the EU (EUDAMED), Class II in the US (FDA GUDID). EU and US classes are different schemes, so this is not a disagreement.

Similar devices in this category

Other devices filed under LINEAR STAPLERS FOR VIDEOSURGERY - OTHER.

Other devices from Intuitive Surgical, Inc.

See all devices from Intuitive Surgical, Inc. →

Properties

  • Sterilise before use
  • Contains CMR substances (carcinogenic, mutagenic or toxic to reproduction)
  • Prescription use
  • Lot / batch number
  • Manufacturing date
Other properties (20, all answered No)