Skip to content

Enpath Deflectable Catheter

by Enpath Medical Incorporated 510(k)

Active United States

510(k) K043489

Regulatory status

as of 29 Oct 2026

RegionAuthorizationNumberDateStatus
United StatesFDA 510(k)SESE Substantially Equivalent · Traditional NumberK043489 Date19 May 2005ClearedStatusActive

Coverage

1 jurisdiction

United States of America
  • Active
  • Expiring
  • Withdrawn
  • Listed
  • Not covered

United States — FDA 510(k) Active

Recalls

2 recalls · FDA

    • Terminated Z-0118-2007 FDA Unclassified (pre-2012) initiated 27 Jun 2006

    Enpath has become aware that some of the Enpath Medical 8F & 9F Steerable sheath devices may have a non-conforming flushport bond that causes a leak where air may enter the flush port chamber.

    • Terminated Z-1573-05 FDA Unclassified (pre-2012) initiated 31 Aug 2005 on shared 510(k) K043489

    Enpath has become aware that some of the Enpath Medical 8 Fr Steerable sheath devices may have non-conforming hemostasis valves in that the valve can leak or the valve was observed to have a small hole.

Recalls on a shared submission name another device cleared under the same FDA submission as this one.