Enzyme Immunoassay Ana Screening Test Kit
510(k) K954723
Active
Valid until
No expiry
FDA recalls
None ongoingDocuments
Trust
Unverified
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK954723 | Date15 Dec 1995 | StatusActive |
Recalls
-
- Terminated Z-0612-2011
Product mislabeled-- the ANA cutoff Control and Prostiva controls were mislabeled. Use of the incorrect controls will cause run failure - patient results cannot be reported because controls will not pass..
-
- Terminated Z-0540-2011
- Terminated Z-0539-2011
The ANA Screening kit generates a high percentage of false positive results, associated with a low check control reading.
-
- Terminated Z-1478-2011
Inaccurate results-- Consumer complaints of increased positivity rates with the ANA screening kit.
-
- Terminated Z-1479-2011
The firm received customer complaints of increased positivity rates associated with the ANA Screening kit due to a decline in stability.
-
- Terminated Z-0613-2011
False positive test results caused by decrease in the OD values of ANA Cutoff Controls and ANA Positive Controls. The increased positivity rates were associated with the Meridian ANA Screening kit. False positive ANA values could result in misdiagnosis or mistreatment .
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- 510(k) Number
- FDA Product Code
- LKJ
- Device Class
- Class II
- Regulation
- 21 CFR Part 866.5100
- Advisory Committee
- Immunology
- FDA Decision
- SESE Substantially Equivalent