Skip to content

Eon Mini™ 3719

by Advanced Neuromodulation Systems

Eon Mini™ 3719 is a Class III device made by Advanced Neuromodulation Systems, approved under FDA PMA P010032 on 21 Nov 2001 and listed in FDA GUDID since 22 Oct 2014 (UDI-DI 05414734402613), active (FDA PMA), not in commercial distribution (GUDID) as of 23 Oct 2026. 14 FDA recalls are linked to it.

Active United States. Also listed in GUDID (US), not in commercial distribution.

PMA

Valid until

No expiry

PMA approval is not time-limited

FDA recalls

8 ongoing

This device is named in the recall

Documents

PMA P010032

FDA submission, from GUDID

Trust

Unverified

Source GUDID, synced 23 Oct 2026.

Regulatory status

as of registry sync 23 Oct 2026

RegionAuthorizationNumberDateStatus
United StatesGUDID listing Number05414734402613 UDI-DI Date22 Oct 2014ListedStatusListed GUDID: not in commercial distribution
United StatesFDA PMAAPPR from GUDID Approved NumberP010032 Date21 Nov 2001DecidedStatusActive

Official records

Across registries

2 registries list this device

  • FDA PMA P010032 decided 21 Nov 2001 Active
  • FDA GUDID UDI-DI 05414734402613 listed 22 Oct 2014 Listed

Recalls

14 recalls · 8 ongoing · FDA

    • Ongoing Z-0458-2026 FDA Class II initiated 6 Oct 2025

    There is a potential issue affecting Eterna Spinal Cord Stimulation (SCS) Implantable Pulse Generators (IPGs), Model 32400 and Liberta RC Deep Brain Stimulation (DBS) Implantable Pulse Generators (IPGs), Model 62400 . These devicess may experience a loss of communication with the Clinician Programmer (CP) and/or the Patient Controller (PC).

  • 7 recalls, same reason

    • Ongoing Z-2508-2023 FDA Class I initiated 18 Jul 2023
    • Ongoing Z-2511-2023 FDA Class I initiated 18 Jul 2023
    • Ongoing Z-2509-2023 FDA Class I initiated 18 Jul 2023
    • Ongoing Z-2510-2023 FDA Class I initiated 18 Jul 2023
    • Ongoing Z-2507-2023 FDA Class I initiated 18 Jul 2023
    • Ongoing Z-2505-2023 FDA Class I initiated 18 Jul 2023
    • Ongoing Z-2506-2023 FDA Class I initiated 18 Jul 2023

    Firm has received complaints from patients who are unable to exit MRI (Magnetic Resonance Imaging) mode on their IPGs (Implantable Pulse Generators).

  • 2 recalls, same reason

    • Terminated Z-2692-2017 FDA Class II initiated 2 Jun 2017
    • Terminated Z-2691-2017 FDA Class II initiated 2 Jun 2017

    The firm discovered a risk of loss of Neuromodulation therapy due to exposure of the IPG to monopolar electrosurgical devices (ESU) during medical procedures.

    • Terminated Z-1785-2013 FDA Class II initiated 30 Nov 2012

    There is issues of warmth or heating at the implant site during charging for the Eon Mini implantable pulse generators (IPGs). In the July 2012 letter St. Jude Medical informed customers that they would be implementing design improvements to the charger to address possible increased energy dissipation when the charger is misaligned with the IPG or the IPG is implanted too near the surface of the

    • Terminated Z-0025-2013 FDA Class II initiated 4 Sep 2012

    Eon Mini IPGs manufactured in April 2012 could potentially exhibit a sudden, brief surge in stimulation that would be felt by the patient as uncomfortable or painful. SJMs routine testing, monitoring, and analysis of manufacturing processes identified three lots of Eon Mini IPGs that the internal battery has the potential to come into contact with the internal microcontroller board.

    • Terminated Z-1738-2013 FDA Class II initiated 26 Jul 2012

    St. Jude Medical has informed your doctor that a number of cases have been reported in which discomfort associated with heating occurred at the device site while patients were using the Charging System to charge their spinal cord stimulator. In three cases, patients received a burn to the skin (one second-degree and two first-degree burns) at the implant site. As of June 30, 2012, St. Jude Medica

    • Terminated Z-3091-2011 FDA Unclassified (pre-2012) initiated 2 Jun 2009

    Product programmed with unapproved software was distributed.

Timeline

16 dated events

  1. Recall initiated FDA recalls Z-2692-2017
  2. Recall initiated FDA recalls Z-2691-2017
  3. Recall initiated FDA recalls Z-2508-2023
  4. Recall initiated FDA recalls Z-2511-2023
  5. Recall initiated FDA recalls Z-2509-2023
  6. Recall initiated FDA recalls Z-2510-2023
  7. Recall initiated FDA recalls Z-2507-2023
  8. Recall initiated FDA recalls Z-2505-2023
  9. Recall initiated FDA recalls Z-2506-2023
  10. Recall initiated FDA recalls Z-0458-2026

6 earlier events are in the sections above.

Coverage

1 jurisdiction

United States of America
  • Active
  • Expiring
  • Withdrawn
  • Listed
  • Not covered

United States — FDA PMA Active

Specification

GUDID

Trade name
Eon Mini™
Generic name
Noninvasive device battery charger
Model / Ref / Catalog
3719
Primary DI / UDI-DI
05414734402613
PMA number
FDA product code
LGW
GMDN term
Noninvasive device battery charger
Device class
Class III
GUDID distribution status
Not in Commercial Distribution

Other devices from Advanced Neuromodulation Systems

  • MN13650

    Class III
  • 1210 Open recall

    Class III
  • Infinity™ 6884 Open recall

    Class III
  • Proclaim™ 3884 Open recall

    Class III

See all devices from Advanced Neuromodulation Systems →

Properties

  • Prescription use
  • Lot / batch number
  • Expiration date
  • Manufacturing date
Other properties (11, all answered No)