Epicor Ablation System
by St Jude Medical 510(k)
Active
510(k) K080292
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK080292 | Date25 Jul 2008 | StatusActive |
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Recalls
-
- Terminated Z-1780-2010
The firm states that overheating of cardiac tissue is caused by inadequate cooling. The potential exists when the UltraWand is run using off-label cooling methods (gravity saline flow instead of pump) that cooling flow to one or both cells can be decreased based on pressure applied by the physician to the device. To reflect this information to the user, the Atrial Fibrillation Division (AFD) has
-
- Terminated Z-0946-2009
Firm has received multiple reports of difficulties connecting the product and/or the presence of error code messages occurring with the use of the connecting cable and the Ablation Control System unit.
Specification
- 510(k) Number
- FDA Product Code
- OCL
- Device Class
- Class II
- Regulation
- 21 CFR Part 878.4400
- Advisory Committee
- General, Plastic Surgery
- FDA Decision
- SESE Substantially Equivalent